FLUID DYNAMIC ANALYSIS USING FLOW VISUALIZATION OF THE BAYLOR/NASA IMPLANTABLE AXIAL-FLOW BLOOD PUMP FOR DESIGN IMPROVEMENT

被引:24
作者
WERNICKE, JT
MEIER, D
MIZUGUCHI, K
DAMM, G
ABER, G
BENKOWSKI, R
NOSE, Y
NOON, GP
DEBAKEY, ME
机构
[1] UNIV ROSTOCK,INST BIOMED ENGN & MED INFORMAT,O-2500 ROSTOCK,GERMANY
[2] NASA,LYNDON B JOHNSON SPACE CTR,HOUSTON,TX 77058
关键词
FLOW VISUALIZATION; FLUID DYNAMIC ANALYSIS; AXIAL FLOW VENTRICULAR ASSIST; HEMOLYSIS;
D O I
10.1111/j.1525-1594.1995.tb02306.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Baylor/NASA Axial Blood Flow Pump has been developed for use as an implantable left ventricular assist device (LVAD). The pump is intended as an assist device for either pulmonary or systemic circulatory support for more than 3-months' duration. To date the pump provides acceptable results in terms of thrombus formation and hemolysis (IH of 0.018 g/100 L). A fluid dynamics analysis using flow visualization was performed to investigate the flow fields and to determine areas within the pump that could be improved. These studies focused upon the inflow area in front of the pump. A prototype axial flow pump assembly was constructed to facilitate the flow visualization studies. Particle image tracking velocimetry techniques were used to measure Amberlite particles suspended in a blood analog fluid composed of 63% water and 37% glycerin. This method used a pulsed (612 Hz) laser light to determine flow velocity profiles, shear stress, Reynolds numbers, and stagnant areas within the axial pump. These studies showed that the flow straightener (a vaned assembly in the pump inflow) reduced Reynolds numbers from 4,640 to 2,540 (at 8.5 L/min) and that the flow straightener exacerbates a discontinuity found between it and the impeller. Within the inflow area, a maximum of 80 N/m(2) shear stress was measured, which is well below published blood damage thresholds. Design variations were investigated resulting in a smoother flow transition between flow straightener and impeller. These variations must be investigated further to establish a correlation with hemolysis and thrombus formation.
引用
收藏
页码:161 / 177
页数:17
相关论文
共 15 条
[1]  
Antaki J F, 1993, ASAIO J, V39, pM231
[2]  
DAUGHTERY R, 1985, FLUID MECHANICS ENG
[3]  
FOX R, INTRO FLUID MECHANIC
[4]  
FRAZIER O, 1989, 9TH ANN M SCI SESS I
[5]  
KERRIGAN J, 1993, ASAIO J, V39, P639
[6]  
LEVERETT LB, 1972, BIOPHYS J, P12
[7]  
Merzkirch W., 1987, FLOW VISUALIZATION
[8]   DEVELOPMENT OF THE BAYLOR NASA AXIAL-FLOW VENTRICULAR ASSIST DEVICE - IN-VITRO PERFORMANCE AND SYSTEMATIC HEMOLYSIS TEST-RESULTS [J].
MIZUGUCHI, K ;
DAMM, GA ;
BOZEMAN, RJ ;
AKKERMAN, JW ;
ABER, GS ;
SVEJKOVSKY, PA ;
BACAK, JW ;
ORIME, Y ;
TAKATANI, S ;
NOSE, Y ;
NOON, GP ;
DEBAKEY, ME .
ARTIFICIAL ORGANS, 1994, 18 (01) :32-43
[9]  
NAITO K, 1994, ISAIO J, V18, P7
[10]   BIO-MEDICUS VENTRICULAR ASSISTANCE [J].
NOON, GP .
ANNALS OF THORACIC SURGERY, 1991, 52 (02) :180-181