AUTOMATIC TORQUE MAGNETOMETER FOR VACUUM-TO-HIGH-PRESSURE HYDROGEN ENVIRONMENTS

被引:3
作者
LARSEN, JW
LIVESAY, BR
机构
[1] Engineering Experiment Station, Georgia Institute of Technology, Atlanta
关键词
D O I
10.1063/1.1135697
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An automatic torque magnetometer has been developed for use in high-pressure hydeogen. It will contain pressures ranging from vacuum to 200 atm of hydrogen gas at sample temperatures greater than 400 °C. This magnetometer, which uses an optical lever postion sensor and a restoring force technique has an operating range of 2.0×103 dyn cm to l.6×10-4 dyn cm. An accompanying digital data collection system extends the sensitivity to 1×10-5 dyn cm as well as increasing the data handling capacity of the system. The magnetic properties of thin films in high-temperature and high-pressure hydrogen environments can be studied using this instruments.
引用
收藏
页码:1285 / 1290
页数:6
相关论文
共 24 条
[1]   COMPUTER-CONTROLLED TORQUE MAGNETOMETER FOR AUTOMATIC-DETERMINATION OF CRYSTAL ORIENTATION [J].
ABE, K ;
CHIKAZUMI, S .
JAPANESE JOURNAL OF APPLIED PHYSICS, 1976, 15 (04) :619-625
[2]   MAGNETIC ANISOTROPY IN SINGLE-CRYSTAL THIN FILMS [J].
BOYD, EL .
IBM JOURNAL OF RESEARCH AND DEVELOPMENT, 1960, 4 (02) :116-129
[3]   USE OF A COMMERCIAL ELECTRIC MICROBALANCE FOR HIGHLY SENSITIVE MAGNETIC TORQUE MEASUREMENTS [J].
BRANSKY, J ;
HIRSCH, AA ;
BRANSKY, I .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1968, 1 (07) :790-&
[4]   DETERMINATION OF MAGNETIC-ANISOTROPY CONSTANTS FROM TORQUE CURVES [J].
BURD, J ;
HUQ, M ;
LEE, EW .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1977, 5 (02) :135-141
[5]   AUTOMATIC RECORDING TORSIONAL MAGNETIC SUSCEPTIBILITY BALANCE [J].
CROFT, GT ;
DONAHOE, FJ ;
LOVE, WF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1955, 26 (04) :360-363
[6]   ELECTROSTATIC TORQUEMETER FOR MAGNETIC-FILM MEASUREMENTS [J].
DEY, SK .
JAPANESE JOURNAL OF APPLIED PHYSICS, 1972, 11 (11) :1628-&
[8]   A DIGITAL VACUUM TORQUE MAGNETOMETER FOR TEMPERATURE RANGE 300-1000 DEGREES K [J].
FLETCHER, EJ ;
DESA, A ;
OREILLY, W ;
BANERJEE, SK .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1969, 2 (04) :311-&
[9]   AN AUTOMATIC COMPENSATED TORQUE BALANCE FOR MEASUREMENT OF MAGNETIC ANISOTROPY IN BROAD TEMPERATURE RANGE FROM HELIUM TEMPERATURE UPWARDS [J].
GERBER, R ;
VILIM, F .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1968, 1 (04) :389-&
[10]   HIGH SENSITIVITY TORSION MAGNETOMETERS FOR OLIGATOMIC FILMS [J].
GRADMANN, U ;
KUMMERLE, W ;
THAM, R .
APPLIED PHYSICS, 1976, 10 (03) :219-225