APPARENT AND ACTUAL C-14 RETENTION IN THE SLOWER TURNOVER BICARBONATE POOL IN MAN WHEN USING LIQUID SCINTILLATION-COUNTING

被引:6
作者
CLAGUE, MB [1 ]
KEIR, MJ [1 ]
CLAYTON, CB [1 ]
机构
[1] ROYAL VICTORIA INFIRM,DEPT MED PHYS,NEWCASTLE TYNE NE1 4LP,TYNE & WEAR,ENGLAND
来源
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES | 1979年 / 30卷 / 10期
关键词
D O I
10.1016/0020-708X(79)90082-6
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The use of liquid scintillation counting to determine 14CO2 expiration without correct calibration of the apparatus suggests that about 25% of the label is retained within the slower turnover bicarbonate pool. Calibration reduces this figure to 13% and is in agreement with the figure obtained using a calibrated ionisatin chamber. The discrepancy is due to reduction in the specific radioactivity in the vial, the mechanism involved being unknown, but it may be a characteristics of certain liquid scintillators under certain conditions. © 1979.