COMPARISON OF 4 TECHNIQUES FOR RECOGNITION OF VENTRICULAR-FIBRILLATION FROM THE SURFACE ECG

被引:52
作者
CLAYTON, RH [1 ]
MURRAY, A [1 ]
CAMPBELL, RWF [1 ]
机构
[1] FREEMAN RD HOSP, DEPT ACAD CARDIOL, NEWCASTLE UPON TYNE NE7 7DN, TYNE & WEAR, ENGLAND
关键词
DEFIBRILLATION; DETECTION ALGORITHM; ECG MONITORING; VENTRICULAR FIBRILLATION;
D O I
10.1007/BF02446668
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Four ventricular fibrillation (VF) detection techniques were assessed using recordings of VF to evaluate sensitivity and VF-like recordings to evaluate specificity. The recordings were obtained from Coronary Care Unit patients. The techniques were: threshold crossing intervals (TCI); peaks in the autocorrelation function (ACF); signal content outside the mean frequency (VF-filter); and signal spectrum shape (spectrum). Using 70 extracts, each 4 s long, from VF recordings, the VF filter achieved a sensitivity of 77 per cent, the ACF, TCI and spectrum algorithms had sensitivities of 67, 53 and 46 per cent, respectively. Susceptibility to false alarms was assessed using 40 extracts from VF-like recordings. The TCI algorithm was the most specific (93 per cent), while the spectrum, VF filter and ACF algorithms had specificities of 72, 55 and 38 per cent, respectively. The TCI algorithm achieved overall sensitivity of 93 per cent and specificity of 60 per cent. The spectrum, VF filter and ACF algorithms had overall sensitivities of 80, 93 and 87 per cent, and overall specificities of 60, 20 and 0 per cent, respectively.
引用
收藏
页码:111 / 117
页数:7
相关论文
共 19 条
[1]  
AUBERT AE, 1982, COMPUT CARDIOL, P477
[2]   ALGORITHMIC SEQUENTIAL DECISION-MAKING IN THE FREQUENCY-DOMAIN FOR LIFE THREATENING VENTRICULAR ARRHYTHMIAS AND IMITATIVE ARTIFACTS - A DIAGNOSTIC SYSTEM [J].
BARRO, S ;
RUIZ, R ;
CABELLO, D ;
MIRA, J .
JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (04) :320-328
[3]   BIO-MEDICAL SIGNAL-PROCESSING (IN 4 PARTS) .1. TIME-DOMAIN METHODS [J].
CHALLIS, RE ;
KITNEY, RI .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (06) :509-524
[4]   VENTRICULAR-FIBRILLATION DETECTION BY A REGRESSION TEST ON THE AUTOCORRELATION FUNCTION [J].
CHEN, S ;
THAKOR, NV ;
MOWER, MM .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1987, 25 (03) :241-249
[5]  
CLAYTON RH, 1992, 1991 P COMP CARD, P685
[6]  
CLAYTON RH, 1991, 1990 P COMP CARD, P515
[7]  
FORSTER FK, 1982, COMPUT CARDIOL, P245
[8]  
Herbschleb J. N., 1979, Computers in Cardiology, P49
[9]  
IDEKER RE, 1990, ANN NY ACAD SCI, V591, P208
[10]   CLINICAL-EXPERIENCE WITH 3 DIFFERENT DEFIBRILLATORS FOR RESUSCITATION OF OUT OF HOSPITAL CARDIAC-ARREST [J].
JAKOBSSON, J ;
REHNQVIST, N ;
NYQVIST, O .
RESUSCITATION, 1990, 19 (02) :167-173