DETERMINATION OF BROMINE IN AIR BY ION MOBILITY SPECTROMETRY

被引:19
作者
KARPAS, Z
POLLEVOY, Y
MELLOUL, S
机构
[1] Department of Analytical Chemistry, Nuclear Research Centre, Negev, Beer-Sheva, 84190
关键词
AIR; BROMINE; ION MOBILITY SPECTROMETRY;
D O I
10.1016/S0003-2670(00)83025-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An ion mobility spectrometer (IMS) was used to study the negative ion chemistry of bromine and monitor its concentration in a series of laboratory experiments and in the ambient air of a chemical production facility. Under the conditions of the IMS, three negative ions, Br-, Br2- and Br3-, were produced from bromine. The relative abundance of Br3- increased with the bromine concentration, whereas those of Br- and Br2- decreased owing to the ion-molecule reactions they undergo with bromine. The presence of chlorine, other bromides and chlorides did not adversely affect the determination of bromine. In all field measurements, the maximum bromine levels measured were below 30 ppb, well below its 8-h time-weighted average threshold limit value (TWA-TLV) of 100 ppb. The rapid response of the IMS was evident from the effect of changes in wind direction or velocity on the measured concentrations.
引用
收藏
页码:503 / 507
页数:5
相关论文