MEASURING CARBON-DIOXIDE TENSION IN SALINE AND ALTERNATIVE SOLUTIONS - QUANTIFICATION OF BIAS AND PRECISION IN 2 BLOOD-GAS ANALYZERS

被引:73
作者
RIDDINGTON, D
VENKATESH, KB
CLUTTONBROCK, T
BION, J
机构
[1] Department of Anaesthetics and Intensive Care, Queen Elizabeth Hospital, Edgbaston, Birmingham
关键词
CARBON DIOXIDE; PARTIAL PRESSURE; TONOMETRY; HYDROGEN ION CONCENTRATION; BLOOD GAS ANALYSIS; PH; INTRAMUCOSAL; CRITICAL ILLNESS;
D O I
10.1097/00003246-199401000-00019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To quantify the precision and bias for two standard blood gas analyzers when measuring PCO2 in five different test solutions (blood, normal saline, human albumin solution, succinylated gelatin 4%, and Hartmann's solution). Design: A comparative laboratory study of two blood gas analyzers measuring known PCO2 in test solutions prepared with a thin film tonometer. Measurements: The PCO2 was determined simultaneously in both blood gas analyzers. The precision and bias for both machines were calculated from the measured PCO2. Results: Analyzer 1 showed a negative bias of between 66.4% and 39.2% when measuring PCO2 in saline. Under the same conditions, analyzer 2 showed a negative bias of between 17.7% and 17.9% for PCO2 in the mid- and high ranges of PCO2, but a positive bias in the low range. Both machines measured PCO2 with satisfactory bias and precision in blood. The use of succinylated gelatin 4% as a test solution improved the bias and precision of both machines. Conclusions: Because of the large bias and precision, neither of these machines is suitable for measuring PCO2 in saline solution. The use of succinylated gelatin 4% as a test solution improves the bias and precision of both machines and we favor this solution for nasogastric tonometry. Even with succinylated gelatin 4%, there is a large negative bias with analyzer 1 and our preference is to use analyzer 2. We recommend that other workers review critically the performance of blood gas electrode systems when measuring PCO2 in saline solution. Because of varying performance of ''standard'' machines, we also recommend that each institution determine its own reference range for intramucosal pH.
引用
收藏
页码:96 / 100
页数:5
相关论文
共 11 条
[1]   EVALUATION OF A NEW THIN-FILM TONOMETER [J].
CHALMERS, C ;
BIRD, BD ;
WHITWAM, JG .
BRITISH JOURNAL OF ANAESTHESIA, 1974, 46 (04) :253-259
[2]   GASTRIC-MUCOSAL PH AS A PROGNOSTIC INDEX OF MORTALITY IN CRITICALLY ILL PATIENTS [J].
DOGLIO, GR ;
PUSAJO, JF ;
EGURROLA, MA ;
BONFIGLI, GC ;
PARRA, C ;
VETERE, L ;
HERNANDEZ, MS ;
FERNANDEZ, S ;
PALIZAS, F ;
GUTIERREZ, G .
CRITICAL CARE MEDICINE, 1991, 19 (08) :1037-1040
[3]  
Fiddian-Green R G, 1992, Intensive Care World, V9, P60
[5]   PREDICTIVE VALUE OF THE STOMACH WALL PH FOR COMPLICATIONS AFTER CARDIAC OPERATIONS - COMPARISON WITH OTHER MONITORING [J].
FIDDIANGREEN, RG ;
BAKER, S .
CRITICAL CARE MEDICINE, 1987, 15 (02) :153-156
[6]  
FIDDIANGREEN RG, 1986, ARCH SURG-CHICAGO, V121, P654
[7]   ADEQUACY OF TISSUE OXYGENATION IN INTACT DOG INTESTINE [J].
GRUM, CM ;
FIDDIANGREEN, RG ;
PITTENGER, GL ;
GRANT, BJB ;
ROTHMAN, ED ;
DANTZKER, DR .
JOURNAL OF APPLIED PHYSIOLOGY, 1984, 56 (04) :1065-1067
[8]   GASTRIC INTRAMUCOSAL PH AS A THERAPEUTIC INDEX OF TISSUE OXYGENATION IN CRITICALLY ILL PATIENTS [J].
GUTIERREZ, G ;
PALIZAS, F ;
DOGLIO, G ;
WAINSZTEIN, N ;
GALLESIO, A ;
PACIN, J ;
DUBIN, A ;
SCHIAVI, E ;
JORGE, M ;
PUSAJO, J ;
KLEIN, F ;
ROMAN, ES ;
DORFMAN, B ;
SHOTTLENDER, J ;
GINIGER, R .
LANCET, 1992, 339 (8787) :195-199
[9]   PROGNOSTIC VALUE OF GASTRIC INTRAMURAL PH IN SURGICAL INTENSIVE-CARE PATIENTS [J].
GYS, T ;
HUBENS, A ;
NEELS, H ;
LAUWERS, LF ;
PEETERS, R .
CRITICAL CARE MEDICINE, 1988, 16 (12) :1222-1224
[10]  
USERS HDB, V1