ANAPHYLACTOID REACTIONS ASSOCIATED WITH REUSE OF HOLLOW-FIBER HEMODIALYZERS AND ACE INHIBITORS

被引:64
作者
PEGUES, DA
BECKSAGUE, CM
WOOLLEN, SW
GREENSPAN, B
BURNS, SM
BLAND, LA
ARDUINO, MJ
FAVERO, MS
MACKOW, RC
JARVIS, WR
机构
[1] CTR DIS CONTROL,NATL CTR INFECT DIS,HOSP INFECT PROGRAM,MAILSTOP A07,ATLANTA,GA 30333
[2] US FDA,FAIRFAX,VA
[3] CONTINENTAL DIALYSIS SPRINGFIELD FAIRFAX,SPRINGFIELD,VA
关键词
D O I
10.1038/ki.1992.409
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
From July 18 through November 27, 1989, 12 anaphylactoid reactions (ARs) occurred in 10 patients at a hemodialysis center in Virginia. One patient required hospitalization; no patients died. ARs occurred within minutes of initiating dialysis and were characterized by peripheral numbness and tingling, laryngeal edema or angioedema, facial or generalized sensation of warmth, and/or nausea or vomiting. All 12 ARs occurred with dialyzers that had been reprocessed with an automated reprocessing system. A cohort study, including all patients undergoing dialysis sessions on the six days when an AR occurred, showed that the patients who experienced ARs were significantly more likely than patients who did not to be treated with angiotensin-converting enzyme (ACE) inhibitors (7/10 vs. 3/33; relative risk = 7.9; 95% confidence interval = 2.5 to 25.2) and to have been exposed to reused dialyzers rather than to new dialyzers (12/70 sessions vs. 0/31; P = 0.016). In those sessions using a reused dialyzer, the mean number of dialyzer uses in case-sessions was significantly higher than for noncase-sessions (10.3 vs. 6.2; P = 0.016). After reuse of dialyzers was discontinued at the center, no further ARs occurred, despite the continued administration of ACE inhibitors. This is the first report of an outbreak of ARs associated exclusively with reused dialyzers. We hypothesize that interactions between a dialyzer that has been repeatedly reprocessed and reused, blood, and additional factors, such as ACE inhibitors, increased the risk of developing ARs.
引用
收藏
页码:1232 / 1237
页数:6
相关论文
共 33 条
[1]   REUSE OF HEMODIALYZERS - RESULTS OF NATIONWIDE SURVEILLANCE FOR ADVERSE-EFFECTS [J].
ALTER, MJ ;
FAVERO, MS ;
MILLER, JK ;
COLEMAN, PJ ;
BLAND, LA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 260 (14) :2073-2076
[2]  
ALTER MJ, 1991, T AM SOC ART INT ORG, V37, P97
[3]  
ALVAREZLARA MA, 1991, LANCET, V337, P370
[4]   HEMODIALYSIS LEUKOPENIA AND COMPLEMENT FUNCTION WITH DIFFERENT DIALYZERS [J].
AMADORI, A ;
CANDI, P ;
SASDELLI, M ;
MASSAI, G ;
FAVILLA, S ;
PASSALEVA, A ;
RICCI, M .
KIDNEY INTERNATIONAL, 1983, 24 (06) :775-781
[5]  
[Anonymous], 1989, STANDARD METHODS EXA
[6]   ANAPHYLAXIS [J].
BOCHNER, BS ;
LICHTENSTEIN, LM .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (25) :1785-1790
[7]  
BOK DV, 1980, P DIAL TRANSPLANT FO, V19, P92
[8]   COMPLEMENT ACTIVATION DURING HEMODIALYSIS - CLINICAL OBSERVATIONS, PROPOSED MECHANISMS, AND THEORETICAL IMPLICATIONS [J].
CHENOWETH, DE .
ARTIFICIAL ORGANS, 1984, 8 (03) :281-287
[9]   ANAPHYLATOXIN FORMATION DURING HEMODIALYSIS - COMPARISON OF NEW AND RE-USED DIALYZERS [J].
CHENOWETH, DE ;
CHEUNG, AK ;
WARD, DM ;
HENDERSON, LW .
KIDNEY INTERNATIONAL, 1983, 24 (06) :770-774
[10]   ANAPHYLATOXIN FORMATION DURING HEMODIALYSIS - EFFECTS OF DIFFERENT DIALYZER MEMBRANES [J].
CHENOWETH, DE ;
CHEUNG, AK ;
HENDERSON, LW .
KIDNEY INTERNATIONAL, 1983, 24 (06) :764-769