DOSIMETRY OF 6-MV X-RAY-BEAM PENUMBRA

被引:45
作者
METCALFE, P [1 ]
KRON, T [1 ]
ELLIOTT, A [1 ]
WONG, T [1 ]
HOBAN, P [1 ]
机构
[1] ROYAL ADELAIDE HOSP,ADELAIDE,SA 5000,AUSTRALIA
关键词
D O I
10.1118/1.597107
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The measurement of x-ray beam dose profiles in the penumbral region, using silicon diode, ionization chamber, TLD, and film dosimetry, has been investigated for a 6-MV beam defined by independent collimators. Penumbral width (80%-20%) at d(max), as measured by diode, film, and TLD was found to be 3.6, 3.6, and 3.4 mm, respectively. These results reflect the relative sensitive widths of each of the measurement systems (2.5, 2.0, and 1.0 mm, respectively). An empirical forming function was used to relate the penumbral shape measured with a finite-sized detector to that which would be measured with a point detector, the width of the point detector penumbra calculated from the diode penumbra is 3.4 mm, indicating that the TLD rods are a good approximation to a point detector. An alternative method of determining the width of a point detector penumbra is to extrapolate the penumbral widths obtained using two or more detectors of sensitive width. With this method, using Farmer and RK ionization chambers, a point detector penumbra width of 3.1 mm is obtained. An EGS4 Monte Carlo simulation, where a point source was assumed, gave a penumbral width of 2.8 mm. Negligible differences between the penumbra of beams defined by symmetric and asymmetric collimators was observed.
引用
收藏
页码:1439 / 1445
页数:7
相关论文
共 23 条
[1]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[2]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[3]   ANALYSIS OF PHYSICAL PARAMETERS ASSOCIATED WITH THE MEASUREMENT OF HIGH-ENERGY X-RAY PENUMBRA [J].
DAWSON, DJ ;
HARPER, JM ;
AKINRADEWO, AC .
MEDICAL PHYSICS, 1984, 11 (04) :491-497
[4]   PENUMBRAL MEASUREMENTS IN WATER FOR HIGH-ENERGY X-RAYS [J].
DAWSON, DJ ;
SCHROEDER, NJ ;
HOYA, JD .
MEDICAL PHYSICS, 1986, 13 (01) :101-104
[5]  
Gager L D, 1977, Med Phys, V4, P494, DOI 10.1118/1.594348
[6]  
HOBAN PW, 1992, THESIS U WAIKATO NZ
[7]  
JOHNS HE, 1984, PHYSICS RADIOLOGY
[8]  
KAHN FM, 1986, MED PHYS, V13, P936
[9]   AN ALGORITHM FOR LEAST-SQUARES ESTIMATION OF NONLINEAR PARAMETERS [J].
MARQUARDT, DW .
JOURNAL OF THE SOCIETY FOR INDUSTRIAL AND APPLIED MATHEMATICS, 1963, 11 (02) :431-441
[10]   BEAM HARDENING OF 10 MV RADIOTHERAPY X-RAYS - ANALYSIS USING A CONVOLUTION SUPERPOSITION METHOD [J].
METCALFE, PE ;
HOBAN, PW ;
MURRAY, DC ;
ROUND, WH .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (11) :1533-1549