PERFORMANCE OF THIN FOIL SCINTILLATING SCREEN FOR TRANSMISSION ELECTRON-MICROSCOPY

被引:12
作者
FAN, GY [1 ]
DUNKELBERGER, DG [1 ]
ELLISMAN, MH [1 ]
机构
[1] GRANT SCI CORP,GILBERT,SC 29054
关键词
D O I
10.1016/0304-3991(94)90074-4
中图分类号
TH742 [显微镜];
学科分类号
摘要
Scintillating screens made by coating a thin metal foil with a layer of phosphor appear to be attractive for transmission electron microscopy applications. We report here the brightness and resolution in the voltage range of 100-400 kV of such a screen made of a 10 mu m layer of P20 phosphor on a 2 mu m Al foil. Both brightness and resolution are superior to that of a screen made by coating a glass plate with a similar layer of phosphor. An exciting property of such a device is that its resolution improves slightly at higher voltages. This, combined with its excellent resistivity to radiation damage and stability under the electron beam, makes it a good candidate for high-voltage applications.
引用
收藏
页码:7 / 14
页数:8
相关论文
共 13 条
[1]   PERFORMANCE OF ELECTRON IMAGE CONVERTERS WITH YAG SINGLE-CRYSTAL SCREEN AND CCD SENSOR [J].
DABERKOW, I ;
HERRMANN, KH ;
LIU, LB ;
RAU, WD .
ULTRAMICROSCOPY, 1991, 38 (3-4) :215-223
[2]  
Dainty J. C., 1974, IMAGE SCI
[3]  
FAN GY, 1993, J ELECTRON MICROSC, V42, P419
[4]   HIGH-SENSITIVITY LENS-COUPLED SLOW-SCAN CCD CAMERA FOR TRANSMISSION ELECTRON-MICROSCOPY [J].
FAN, GY ;
ELLISMAN, MH .
ULTRAMICROSCOPY, 1993, 52 (01) :21-29
[5]  
Herrmann K.H., 1984, ADV OPTICAL ELECT MI, P1
[6]  
HERRMANN KH, 1992, OPTIK, V92, P48
[7]   ANALYSIS OF ELECTRON IMAGE DETECTION EFFICIENCY OF SLOW-SCAN CCD CAMERAS [J].
ISHIZUKA, K .
ULTRAMICROSCOPY, 1993, 52 (01) :7-20
[8]  
IWANAGA M, 1968, Journal of Electron Microscopy, V17, P203
[9]   PERFORMANCE OF A LOW-NOISE CCD CAMERA ADAPTED TO A TRANSMISSION ELECTRON-MICROSCOPE [J].
KUJAWA, S ;
KRAHL, D .
ULTRAMICROSCOPY, 1992, 46 (1-4) :395-403
[10]  
PAWLEY J, 1980, 38TH P ANN EMSA M, P8