QUANTIFICATION OF THE EFFECTS OF RESPIRATION AND PARALLAX ON INFERIOR VENA-CAVAL FILTER POSITION

被引:6
作者
WOLFER, GK [1 ]
TAYLOR, FC [1 ]
SMITH, DC [1 ]
机构
[1] LOMA LINDA UNIV,MED CTR,DEPT RADIOL,11234 ANDERSON ST,LOMA LINDA,CA 92354
关键词
VENAE-CAVAE; FILTERS;
D O I
10.1016/S1051-0443(94)71502-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: The change in inferior vena caval (IVC) filter position at follow-up relative to the filter position at implantation has been used as a criterion for evaluation and comparison of these devices. Perceived changes in filter position may be due to respiratory movement and/or changes in parallax between the initial and follow-up imaging studies. In this study the authors evaluated and attempted to quantify the effects of respiratory movement and parallax. PATIENTS AND METHODS: After placement of an IVC filter, radiographs of the abdomen were taken at maximum inspiration and maximum expiration in 30 patients. The effect of parallax on apparent filter movement was studied by using a phantom. RESULTS: The average filter movement on inspiration/expiration radiographs (corrected for magnification) was 3.6 mm +/- 2.2. An 8.5-mm maximal change secondary to parallax was seen in the phantom study. CONCLUSION: When follow-up images are obtained, efforts should be made to closely reproduce patient positioning and patient respiration to reduce errors in the interpretation of filter migration.
引用
收藏
页码:357 / 360
页数:4
相关论文
共 13 条
[1]  
Greenfield L J, 1991, J Vasc Interv Radiol, V2, P425, DOI 10.1016/S1051-0443(91)72210-0
[2]   EVOLUTION OF HOOK DESIGN FOR FIXATION OF THE TITANIUM GREENFIELD FILTER [J].
GREENFIELD, LJ ;
CHO, KJ ;
TAUSCHER, JR .
JOURNAL OF VASCULAR SURGERY, 1990, 12 (03) :345-353
[3]  
GREENFIELD LJ, 1981, ARCH SURG-CHICAGO, V116, P1451
[4]  
GREENFIELD LJ, 1991, J VASC SURG, V14, P253
[5]   EXPERIENCE WITH THE GIANTURCO-ROEHM BIRDS NEST VENA-CAVA FILTER [J].
MARTIN, B ;
MARTYAK, TE ;
STOUGHTON, TL ;
COLLAZO, WA ;
PEARL, W .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (17) :1275-1277
[6]   GREENFIELD CAVAL FILTERS - LONG-TERM RADIOGRAPHIC FOLLOW-UP-STUDY [J].
MESSMER, JM ;
GREENFIELD, LJ .
RADIOLOGY, 1985, 156 (03) :613-618
[7]  
Millward S F, 1991, J Vasc Interv Radiol, V2, P429, DOI 10.1016/S1051-0443(91)72212-4
[8]  
Murphy T P, 1991, J Vasc Interv Radiol, V2, P107, DOI 10.1016/S1051-0443(91)72482-2
[9]  
RICO JB, 1988, ANN VASC SURG, V3, P242
[10]   THE BIRDS NEST INFERIOR VENA-CAVA FILTER - PROGRESS REPORT [J].
ROEHM, JOF ;
JOHNSRUDE, IS ;
BARTH, MH ;
GIANTURCO, C .
RADIOLOGY, 1988, 168 (03) :745-749