OPTICAL-DETECTION OF MACROMOLECULAR HEPARIN VIA SELECTIVE COEXTRACTION INTO THIN POLYMERIC FILMS

被引:38
作者
WANG, EJ
MEYERHOFF, ME
YANG, VC
机构
[1] UNIV MICHIGAN, DEPT CHEM, ANN ARBOR, MI 48109 USA
[2] UNIV MICHIGAN, COLL PHARM, ANN ARBOR, MI 48109 USA
关键词
D O I
10.1021/ac00099a007
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Thin plasticized polymer films, poly(vinyl chloride) doped with a specific ion pairing quaternary ammonium compound, tridodecylmethylammonium chloride, and a lipophilic pH indicator, 3-hydroxy-4-(4-nitrophenylazo)phenyl octadeconate, are shown to exhibit significant and analytically useful optical response toward macromolecular heparin. The response mechanism is based on favorable extraction of heparin into the bulk organic film, owing to the specific ion-pairing complexation reaction between the quaternary ammonium species and the polyanion. A simultaneous coextraction of hydrogen ions results in protonation of the pH chromophore and hence a change in the optical absorbance of the polymeric film. When used in a limited volume/fixed exposure (10 min) detection mode, film absorbances change as a function of the initial heparin concentration in the range of 0.2-3.0 units/mL (1.2-18 mu g/mL). The practical measurement response time is controlled by heparin diffusion through the stagnant diffusion layer adjacent to the surface of the film as well as within the bulk of the polymer film and is shown to increase with the molecular weight of the heparin species tested. No optical response to heparin is observed when a strong heparin complexing agent (e.g., protamine) is present in the test solution, suggesting that the polymer film can be used to conveniently monitor heparin-protamine (or other antagonist) titrations. The theory relating to the operation of the sensing film in either the equilibrium or the kinetic mode and the selectivity of the optimized film to heparin relative to small anions are presented.
引用
收藏
页码:522 / 527
页数:6
相关论文
共 22 条
[1]   SELECTIVITY OF ION-SENSITIVE BULK OPTODES [J].
BAKKER, E ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1805-1812
[2]  
CASU B, 1989, ANN NEW YORK ACADEMY, V556
[3]   RESPONSE MECHANISM OF POLYMER MEMBRANE-BASED POTENTIOMETRIC POLYION SENSORS [J].
FU, B ;
BAKKER, E ;
YUN, JH ;
YANG, VC ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1994, 66 (14) :2250-2259
[4]  
GANTZER ML, 1985, Patent No. 153641
[5]   INHIBITION OF LOW-MOLECULAR WEIGHT HEPARIN BY PROTAMINE CHLORIDE INVIVO [J].
HARENBERG, J ;
GNASSO, A ;
DEVRIES, JX ;
ZIMMERMANN, R ;
AUGUSTIN, J .
THROMBOSIS RESEARCH, 1985, 38 (01) :11-20
[6]  
JANATA J, 1990, ANAL CHEM, V62, pR33, DOI 10.1021/ac00211a017
[7]   BISULFITE ADDITION-REACTION AS THE BASIS FOR A HYDROGENSULFITE BULK OPTODE [J].
KURATLI, M ;
BADERTSCHER, M ;
RUSTERHOLZ, B ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1993, 65 (23) :3473-3479
[8]   LEAD-SELECTIVE BULK OPTODES BASED ON NEUTRAL IONOPHORES WITH SUBNANOMOLAR DETECTION LIMITS [J].
LERCHI, M ;
BAKKER, E ;
RUSTERHOLZ, B ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1992, 64 (14) :1534-1540
[9]   BULK OPTODES BASED ON NEUTRAL DITHIOCARBAMATE IONOPHORES WITH HIGH SELECTIVITY AND SENSITIVITY FOR SILVER AND MERCURY CATIONS [J].
LERCHI, M ;
REITTER, E ;
SIMON, W ;
PRETSCH, E ;
CHOWDHURY, DA ;
KAMATA, S .
ANALYTICAL CHEMISTRY, 1994, 66 (10) :1713-1717
[10]  
LINHARDT RJ, 1991, CHEM IND-LONDON, P45