SIZES AND SOURCES OF FIELD PLACEMENT ERROR IN ROUTINE IRRADIATION FOR PROSTATE-CANCER

被引:18
作者
DUNSCOMBE, P [1 ]
LOOSE, S [1 ]
LESZCZYNSKI, K [1 ]
机构
[1] LAURENTIAN UNIV,DEPT PHYS,SUDBURY P3E 2C6,ONTARIO,CANADA
关键词
PORTAL VERIFICATION; PROSTATE CANCER; RADIOTHERAPY QUALITY CONTROL;
D O I
10.1016/0167-8140(93)90100-M
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Discrepancies between prescribed and treated field edges have been measured from 76 film pairs taken on 29 patients being treated for prostate cancer. The distribution of field edge discrepancies is described by a standard deviation of 4.5 mm and has an average absolute value of 3.5 mm. The observed discrepancies are shown to be attributable to field position errors and hence, in principle, are easily rectifiable.
引用
收藏
页码:174 / 176
页数:3
相关论文
共 4 条
  • [1] ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH
    DENEVE, W
    VANDENHEUVEL, F
    DEBEUKELEER, M
    COGHE, M
    THON, L
    DEROOVER, P
    VANLANCKER, M
    STORME, G
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) : 45 - 54
  • [2] PRECISION OF DETERMINING COMPLIANCE WITH PRESCRIBED FIELDS FROM CONVENTIONAL PORTAL FILMS
    DUNSCOMBE, PB
    FOX, K
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (733) : 48 - 52
  • [3] DUNSCOMBE PB, UNPUB INT J RAD ONCO
  • [4] ACCURACY OF RADIATION-FIELD ALIGNMENT IN CLINICAL-PRACTICE
    RABINOWITZ, I
    BROOMBERG, J
    GOITEIN, M
    MCCARTHY, K
    LEONG, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (10): : 1857 - 1867