MEMBRANE FLUIDITY MEASURED BY FLUORESCENCE POLARIZATION USING AN EPICS V-CELL SORTER
被引:42
作者:
FOX, MH
论文数: 0引用数: 0
h-index: 0
机构:COLORADO STATE UNIV, DEPT RADIOL & RADIAT BIOL, FT COLLINS, CO 80523 USA
FOX, MH
DELOHERY, TM
论文数: 0引用数: 0
h-index: 0
机构:COLORADO STATE UNIV, DEPT RADIOL & RADIAT BIOL, FT COLLINS, CO 80523 USA
DELOHERY, TM
机构:
[1] COLORADO STATE UNIV, DEPT RADIOL & RADIAT BIOL, FT COLLINS, CO 80523 USA
[2] COLUMBIA UNIV, CELL SORTER LAB, NEW YORK, NY 10027 USA
来源:
CYTOMETRY
|
1987年
/
8卷
/
01期
关键词:
D O I:
10.1002/cyto.990080104
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
We have characterized the measurement of fluorescence polarization on single cells using an EPICS V cell sorter. A critical analysis is made of the balancing and calibration of the system. The system is highly linear for polarization measurements. Cellular membranes were labeled with 1,6-diphenyl-1,3,5-hexatriene (DPH) to measure membrane fluidity. Fluorescence polarization histograms had coefficients of variation as low as 7%. Cells labeled with DPH after 24 hr incubation in medium lacking serum showed a significantly higher fluorescence polarization than cells in medium containing serum. The fluorescence polarization measured at 15.degree. C was 0.311 compared to 0.270 at 25.degree. C for cells labeled with DPH, verifying that temperature affects the membrane fluidity as measured by flow cytometry.