THE DETERMINATION OF TOTAL SERUM-CHOLESTEROL BY FLOW-INJECTION ANALYSIS WITH AMPEROMETRIC DETECTION

被引:15
作者
CARPENTER, A
PURDY, WC
机构
[1] Department of Chemistry, McGill University, Montreal, Quebec
基金
加拿大自然科学与工程研究理事会;
关键词
amperometric detection; enzyme reactors; flow-injection analysis; Serum Cholesterol;
D O I
10.1080/00032719008052454
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-injection system is described which employs immobilized cholesterol esterase and cholesterol oxidase for the determination of total serum cholesterol. Hydrogen peroxide generated by the enzymatic reaction is measured directly at a platinum electrode held at +600 mV vs Ag/AgCl. Serum samples are passed through an anion exchange resin to remove possible interferences prior to analysis. The accuracy and precision of this system was evaluated with a Standard Reference Material for serum cholesterol and the quantitative performance was found to conform to recent guidelines put forward by the National Institutes of Health’s National Cholesterol Education Program. © 1990, Taylor & Francis Group, LLC. All rights reserved.