COMPUTERIZED TOMOGRAPHY COMPUTER GRAPHICS APPROACH TO STEREOTAXIC LOCALIZATION

被引:172
作者
BROWN, RA
机构
关键词
D O I
10.3171/jns.1979.50.6.0715
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
This paper describes the use of three-dimensional computer graphics to display structures visible in computerized tomography (CT) scans and to accurately determine optimum stereotaxic probe placement relative to those structures. A prototype Lucite stereotaxic frame designed for use in CT body scanners was fitted with a phantom consisting of small Lucite spheres representing intracranial tumors with diameters of 6 to 19 mm. A series of CT scans was obtained of the frame and phantom together. Edge outlines of the spheres were extracted from each scan in the series. A three-dimensional visual representation of the spheres was obtained by displaying their outlines from all CT scans. A superimposed visual simulation of the stereotaxic frame was adjusted interactively using several analog dials in order to simulate trajectory choice and probe depth before actual surgery. The frame settings and probe depth calculated by the graphical computer were then applied to the actual prototype frame in order to assess the accuracy of this combined CT-computer graphics approach to stereotaxic localization. The results of 22 such experiments and the implications for future clinical use of this new and precise localization technique are discussed.
引用
收藏
页码:715 / 720
页数:6
相关论文
共 8 条
  • [1] STEREOTAXIC COMPUTED TOMOGRAPHY
    BERGSTROM, M
    GREITZ, T
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1976, 127 (01) : 167 - 170
  • [2] CETAS T C, 1978, Medical Physics (Woodbury), V5, P79, DOI 10.1118/1.594414
  • [3] 1975 MEMORIAL AWARD PAPER - IMAGE GENERATION AND DISPLAY TECHNIQUES FOR CT SCAN DATA - THIN TRANSVERSE AND RECONSTRUCTED CORONAL AND SAGITTAL PLANES
    GLENN, WV
    JOHNSTON, RJ
    MORTON, PE
    DWYER, SJ
    [J]. INVESTIGATIVE RADIOLOGY, 1975, 10 (05) : 403 - 416
  • [4] FURTHER INVESTIGATION AND INITIAL CLINICAL USE OF ADVANCED CT DISPLAY CAPABILITY
    GLENN, WV
    JOHNSTON, RJ
    MORTON, PE
    DWYER, SJ
    [J]. INVESTIGATIVE RADIOLOGY, 1975, 10 (05) : 479 - 489
  • [5] POTENTIAL ARTIFACTS ASSOCIATED WITH SCANNING PATTERN OF EMI SCANNER
    GOODENOUGH, DJ
    WEAVER, KE
    DAVIS, DO
    [J]. RADIOLOGY, 1975, 117 (03) : 615 - 619
  • [6] Mundinger F, 1974, Acta Neurochir (Wien), VSuppl 21, P93
  • [7] RHODES ML, 1978, P SAN DIEGO BIOMEDIC, V17, P403
  • [8] SORENSON JA, MED PHYS