AN ELECTROENZYMATIC GLUCOSE SENSOR FABRICATED ON A FLEXIBLE SUBSTRATE

被引:26
作者
MASTROTOTARO, JJ
JOHNSON, KW
MORFF, RJ
LIPSON, D
ANDREW, CC
ALLEN, DJ
机构
[1] Medical Devices and Diagnostics Division, Eli Lilly and Company, Indianapolis
关键词
D O I
10.1016/0925-4005(91)80234-B
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A glucose-oxidase-based electroenzymatic glucose sensor has been developed using thin-/thick-film processing techniques. We believe that this processing scheme will overcome a major impediment to the successful movement of the devices from the research laboratory to the marketplace by removing the difficulty of fabricating large numbers of reproducible and economical sensors. Several hundred sensors have been fabricated and tested in vitro. The sensors respond linearly to glucose, are stable for 72 h, are oxygen independent and have a 90 s response time. The sensors have also responded appropriately during preliminary glucose tolerance tests performed in rabbits.
引用
收藏
页码:139 / 144
页数:6
相关论文
共 8 条
[1]   APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS [J].
ARMOUR, JC ;
LUCISANO, JY ;
MCKEAN, BD ;
GOUGH, DA .
DIABETES, 1990, 39 (12) :1519-1526
[2]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&
[3]   FUNDAMENTALS OF GLUCOSE SENSORS [J].
FISCHER, U .
DIABETIC MEDICINE, 1991, 8 (04) :309-321
[4]   REPRODUCIBLE ELECTRODEPOSITION OF BIOMOLECULES FOR THE FABRICATION OF MINIATURE ELECTROENZYMATIC BIOSENSORS [J].
JOHNSON, KW .
SENSORS AND ACTUATORS B-CHEMICAL, 1991, 5 (1-4) :85-89
[5]  
JOHNSON KW, 1989, 175TH M EL SOC, V89, P921
[6]  
KERNER W, 1988, HORM METAB RES, V20, P8
[7]   INVIVO MOLECULAR SENSING IN DIABETES-MELLITUS - AN IMPLANTABLE GLUCOSE SENSOR WITH DIRECT ELECTRON-TRANSFER [J].
PICKUP, JC ;
SHAW, GW ;
CLAREMONT, DJ .
DIABETOLOGIA, 1989, 32 (03) :213-217
[8]  
SHICHIRI M, 1988, HORM METAB RES, V20, P17