RAPID ASSESSMENT OF ENDOGENOUS CREATININE IN PHYSIOLOGICAL SAMPLES ANALYZED BY REVERSED-PHASE LIQUID-CHROMATOGRAPHY WITH SPECTROPHOTOMETRIC DETECTION

被引:13
作者
KRSTULOVIC, AM [1 ]
BERTANIDZIEDZIC, L [1 ]
CAPORUSSO, JM [1 ]
机构
[1] CUNY MT SINAI SCH MED,DEPT MED,CATECHOLAMINE RES LAB,NEW YORK,NY 10029
关键词
D O I
10.1016/0009-8981(79)90042-1
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
An improved and direct analysis of endogenous creatinine levels using the reversed-phase mode of high performance liquid chromatography (RPLC) is described. The analysis is rapid, the sensitivity of the spectrophotometric detection is in the nanomole range, and the sample preparation requires only filtration to remove the particulate matter. Identification of chromatographic peaks was achieved by means of absorbance ratios, stopped-flow UV spectra and prechromatographic incubation of samples with creatinine amidohydrolase. Reported in this paper are representative chromatograms of samples of amniotic fluid, urine, and cerebrospinal fluid (CSF). The ranges of creatinine concentrations in 10 samples of amniotic fluid, urine and CSF were 0.774-1.850, 99-454.5 and 0.648-1.05 mg/100 ml, respectively. This highly specific and simple assay offers several improvements over previously used methods. © 1979.
引用
收藏
页码:189 / 194
页数:6
相关论文
共 13 条