EFFECT OF ION-LENS TUNING AND FLOW-INJECTION ON NONSPECTROSCOPIC MATRIX INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:38
作者
WANG, JS
SHEN, WL
SHEPPARD, BS
EVANS, EH
CARUSO, JA
FRICKE, FL
机构
[1] UNIV CINCINNATI,DEPT CHEM,CINCINNATI,OH 45221
[2] US FDA,ELEMENTAL ANALYS RES CTR,CINCINNATI,OH 45202
关键词
Flow injection; Inductively coupled plasma mass spectrometry; Ion-lens tuning; Matrix-induced suppression;
D O I
10.1039/ja9900500445
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effects of various matrix constituents, Co, Pb and synthetic ocean water, on analyte signal have been studied. Suppression of the 138Ba + signal was observed in the presence of each matrix, with the Pb matrix having the greatest effect. These matrix-induced suppressions of analyte signal were compensated for in part by the use of In as an internal standard, and by initially tuning the ion lenses for maximum analyte signal in the presence of the matrix, rather than for a simple acidified standard solution. The severity of matrix-induced suppression was compared for flow injection and continuous sample introduction. No significant differences in suppression of analyte signal between the two methods were observed because of the relatively large sample volume used for flow injection. The experimental procedure adopted, continual reference of analyte signals with and without the matrix present, effectively eliminated any bias in the results due to nebuliser and cone blockage.