INJECTION METHOD FOR THE SEQUENTIAL DETERMINATION OF BORON AND SEVERAL METALS IN WASTEWATER SAMPLES BY INDUCTIVELY-COUPLED PLASMA ATOMIC EMISSION SPECTROMETRY

被引:50
作者
BROEKAERT, JAC
LEIS, F
机构
关键词
D O I
10.1016/S0003-2670(01)84229-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Boron and nine metals (Ba, Cd, Cu, Fe, Mn, Mo, Ni, Zn) can be determined in waste waters of medium and high salt concentration by inductively-coupled plasma atomic emission spectrometry. An injection method was applied in order to achieve analysis times of about 1 min per determination, and to allow trace analysis in solutions having salt concentrations up to 20 mg ml-1, when a Meinhard nebulizer is used with a low-power argon plasma. When 500-μl aliquots are injected and the peak heights of the amplified photomultiplier signals are measured, the detection limits for the elements tested range from 0.05 to 0.4 μg ml-1. Matrix effects caused by sodium nitrate (20 mg ml-1) are less than 20%, provided that a peristaltic pump is used for feeding the nebulizer. Matrix effects can be decreased by using simultaneous background measurement. The plasma-injection method is applicable to boron and metals in the range 0.05-20 μg ml-1. © 1979.
引用
收藏
页码:73 / 83
页数:11