HYBRID BIOSENSOR FOR THE DETERMINATION OF SUCROSE

被引:48
作者
BARLIKOVA, A
SVORC, J
MIERTUS, S
机构
[1] Department of Analytical Chemistry, Slovak Technical University, 812 37 Bratislava
关键词
AMPEROMETRY; BIOSENSORS; ENZYME REACTOR; FOODS; SUCROSE; YEAST CELL WALLS;
D O I
10.1016/S0003-2670(00)83056-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Yeast cell walls of Saccharomyces cerevisiae as a source of invertase together with glucose oxidase were used for the construction of a hybrid sucrose sensor. The biocatalytic layer was prepared by co-immobilizing yeast cell walls and glucose oxidase on a nylon network via glutardialdehyde and was fixed to the Clark oxygen electrode. The mutarotation of alpha- to beta-glucose was accelerated by phosphate ions instead of mutarotase. The influence of pH, temperature and optimum composition of the biocatalytic layer was tested. Test samples and real samples were examined and the results were compared with those obtained by means of a glucose electrode. The sensor is stable for more than 2 months and the linear range extends to 1.3 x 10(-3) mol 1(-1).
引用
收藏
页码:83 / 87
页数:5
相关论文
共 20 条
[1]  
BERGMAYER HU, 1984, METHODS ENZYMATIC AN
[2]   MAGNETIC ENZYME MEMBRANES AS ACTIVE ELEMENTS OF ELECTROCHEMICAL SENSORS - LACTOSE, SACCHAROSE, MALTOSE BIENZYME ELECTRODES [J].
CORDONNIER, M ;
LAWNY, F ;
CHAPOT, D ;
THOMAS, D .
FEBS LETTERS, 1975, 59 (02) :263-267
[3]   ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE [J].
GUILBAULT, GG ;
LUBRANO, GJ .
ANALYTICA CHIMICA ACTA, 1973, 64 (03) :439-455
[4]   CHEMICALLY IMMOBILIZED TRI-ENZYME ELECTRODE FOR THE DETERMINATION OF SUCROSE USING FLOW-INJECTION ANALYSIS [J].
HAMID, JA ;
MOODY, GJ ;
THOMAS, JDR .
ANALYST, 1988, 113 (01) :81-85
[5]   THE DEVELOPMENT OF NEW ANALYTICAL SYSTEMS BASED ON BIOCATALYSTS [J].
KULYS, JJ .
ANALYTICAL LETTERS PART B-CLINICAL AND BIOCHEMICAL ANALYSIS, 1981, 14 (06) :377-397
[6]   A BIOSPECIFIC MEMBRANE SENSOR FOR THE DETERMINATION OF SUCROSE [J].
MACHOLAN, L ;
KONECNA, H .
COLLECTION OF CZECHOSLOVAK CHEMICAL COMMUNICATIONS, 1983, 48 (03) :798-804
[7]   SIMULTANEOUS DETERMINATION OF GLUCOSE, FRUCTOSE, AND SUCROSE IN MIXTURES BY AMPEROMETRIC FLOW-INJECTION ANALYSIS WITH IMMOBILIZED ENZYME REACTORS [J].
MATSUMOTO, K ;
KAMIKADO, H ;
MATSUBARA, H ;
OSAJIMA, Y .
ANALYTICAL CHEMISTRY, 1988, 60 (02) :147-151
[8]   MODEL FOR AMPEROMETRIC ENZYME ELECTRODE OBTAINED THROUGH DIGITAL-SIMULATION AND APPLIED TO IMMOBILIZED GLUCOSE OXIDASE SYSTEM [J].
MELL, LD ;
MALOY, JT .
ANALYTICAL CHEMISTRY, 1975, 47 (02) :299-307
[9]   SIMULTANEOUS DETERMINATION OF GLUCOSE AND SUCROSE BY A GLUCOSE-SENSING ENZYME ELECTRODE COMBINED WITH AN INVERTASE-ATTACHED CELL [J].
MIZUTANI, F ;
ASAI, M .
ANALYTICA CHIMICA ACTA, 1990, 236 (02) :245-250
[10]   GALACTOSE DETERMINATION IN AN AUTOMATED FLOW-INJECTION SYSTEM CONTAINING ENZYME REACTORS AND AN ONLINE DIALYZER [J].
OLSSON, B ;
LUNDBACK, H ;
JOHANSSON, G .
ANALYTICA CHIMICA ACTA, 1985, 167 (JAN) :123-136