MAGNETOENCEPHALOGRAPHIC LOCALIZATION OF EPILEPTIC CORTEX - IMPACT ON SURGICAL-TREATMENT

被引:63
作者
PAETAU, R
KAJOLA, M
KARHU, J
NOUSIAINEN, U
PARTANEN, J
TIIHONEN, J
VAPALAHTI, M
HARI, R
机构
[1] KUOPIO UNIV HOSP,DEPT CLIN NEUROPHYSIOL,KUOPIO,FINLAND
[2] KUOPIO UNIV HOSP,DEPT NEUROSURG,KUOPIO,FINLAND
[3] VAAJASALO HOSP,KORTEJOKI,FINLAND
关键词
D O I
10.1002/ana.410320119
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
A 24-channel, planar, superconducting quantum interference device gradiometer, sampling a fourth of the head surface over brain tissue, was used to determine the site of an epileptic focus in a 36-year-old woman with intractable complex partial epilepsy. The other presurgical findings appeared divergent: a large arachnoid cyst over the right parietal convexity, dissimilar interictal electroencephalographic patterns, and several neuropsychological dysfunctions. The equivalent current sources of magnetoencephalographic spikes were in the right posterior temporal region of the cortex, 4 cm apart from the cyst. Surgical exploration of the area pinpointed by magnetoencephalography revealed a pachygyric patch of cortex displaying focal discharges on the electrocorticogram. After resection, a dramatic reduction of seizures occurred. The good agreement between electrocorticography and magnetoencephalography warrants future investigation of multichannel magnetoencephalography as a potential alternative to invasive presurgical recordings.
引用
收藏
页码:106 / 109
页数:4
相关论文
共 13 条
  • [1] MULTICHANNEL SQUID SYSTEMS FOR BRAIN RESEARCH
    AHONEN, AI
    HAMALAINEN, MS
    KAJOLA, MJ
    KNUUTILA, JET
    LOUNASMAA, OV
    SIMOLA, JT
    TESCHE, CD
    VILKMAN, VA
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (02) : 2786 - 2792
  • [2] NEUROMAGNETIC LOCALIZATION OF EPILEPTIFORM SPIKE ACTIVITY IN THE HUMAN-BRAIN
    BARTH, DS
    SUTHERLING, W
    ENGEL, J
    BEATTY, J
    [J]. SCIENCE, 1982, 218 (4575) : 891 - 894
  • [3] RECORDING AND INTERPRETATION OF CEREBRAL MAGNETIC-FIELDS
    HARI, R
    LOUNASMAA, OV
    [J]. SCIENCE, 1989, 244 (4903) : 432 - 436
  • [4] Hari R., 1990, AUDITORY EVOKED MAGN, P222
  • [5] LARGE-AREA LOW-NOISE 7-CHANNEL DC SQUID MAGNETOMETER FOR BRAIN RESEARCH
    KNUUTILA, J
    AHLFORS, S
    AHONEN, A
    HALLSTROM, J
    KAJOLA, M
    LOUNASMAA, OV
    VILKMAN, V
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (11) : 2145 - 2156
  • [6] MODENA I, 1982, CLIN NEUROPHYSIOL, V54, P622
  • [7] STUDY OF FOCAL EPILEPSY BY MULTICHANNEL NEUROMAGNETIC MEASUREMENTS
    RICCI, GB
    ROMANI, GL
    SALUSTRI, C
    PIZZELLA, V
    TORRIOLI, G
    BUONOMO, S
    PERESSON, M
    MODENA, I
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1987, 66 (04): : 358 - 368
  • [8] MAGNETOENCEPHALOGRAPHY AND EPILEPSY RESEARCH
    ROSE, DF
    SMITH, PD
    SATO, S
    [J]. SCIENCE, 1987, 238 (4825) : 329 - 335
  • [9] STEFAN H, 1989, ADV BIOMAGNETISM, P279
  • [10] THE MAGNETIC-FIELD OF COMPLEX PARTIAL SEIZURES AGREES WITH INTRACRANIAL LOCALIZATIONS
    SUTHERLING, WW
    CRANDALL, PH
    ENGEL, J
    DARCEY, TM
    CAHAN, LD
    BARTH, DS