THERMAL-CONDUCTIVITY OF ELECTRICALLY CONDUCTING LIQUIDS BY THE TRANSIENT HOT-WIRE METHOD

被引:51
作者
BLEAZARD, JG [1 ]
TEJA, AS [1 ]
机构
[1] GEORGIA INST TECHNOL,SCH CHEM ENGN,FLUID PROPERTIES RES INST,ATLANTA,GA 30332
关键词
D O I
10.1021/je00020a003
中图分类号
O414.1 [热力学];
学科分类号
摘要
A relative transient hot-wire apparatus for the measurement of the thermal conductivity of electrically conducting liquids is described. The instrument consists of a single glass capillary filled with mercury to act as an insulated hot wire. The resistance change in the wire with respect to time is used to obtain the thermal conductivity of the liquid surrounding the wire with an estimated accuracy of +/-2%. The most significant advant;age of the liquid metal filled glass capillary is its increased temperature limit compared to other types of transient hot-wire cells. The use of the new method at temperatures up to 493 K is shown for several aqueous systems, and new data are reported for propionic acid + water mixtures.
引用
收藏
页码:732 / 737
页数:6
相关论文
共 31 条
[1]  
Abramowitz M., 1965, HDB MATH FUNCTIONS
[2]  
Alloush A., 1982, International Journal of Thermophysics, V3, P225, DOI 10.1007/BF00503318
[3]  
BLEAZARD JG, UNPUB INT J THERMOPH
[4]  
BLEAZARD JG, 1994, AICHE S SER, V90, P23
[5]  
Carslaw H. S., 1986, CONDUCTION HEAT SOLI
[6]  
DIGUILIO RM, 1990, ASHRAE TRAN, V96, P702
[7]   THE THERMAL-CONDUCTIVITY OF THE MOLTEN NANO3-KNO3 EUTECTIC BETWEEN 525-K AND 590-K [J].
DIGUILIO, RM ;
TEJA, AS .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 1992, 13 (04) :575-592
[8]   THEORY OF TRANSIENT HOT-WIRE METHOD FOR MEASURING THERMAL-CONDUCTIVITY [J].
HEALY, JJ ;
DEGROOT, JJ ;
KESTIN, J .
PHYSICA B & C, 1976, 82 (02) :392-408
[9]  
Ho C., 1972, J PHYS CHEM REFERENC, V1, P279, DOI DOI 10.1063/1.3253100
[10]   TRANSIENT METHOD TO MEASURE THE THERMAL-CONDUCTIVITY OF HIGH-TEMPERATURE MELTS USING A LIQUID-METAL PROBE [J].
HOSHI, M ;
OMOTANI, T ;
NAGASHIMA, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (05) :755-758