IMPROVED ACCURACY OF THE LASER DIFFRACTION TECHNIQUE FOR DIAMETER MEASUREMENT OF SMALL FIBERS

被引:26
作者
LI, CT
TIETZ, JV
机构
[1] Dow Corning Corporation, Midland, 48686, Michigan
关键词
D O I
10.1007/BF01129926
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The accuracy of a laser diffraction technique has been improved for small fibre diameter measurement by a calibration with SEM-measured internal standards. The calibration showed that the laser diameter is about 5% larger than the corresponding SEM diameter. After correction for systematic error via accurate calibration, the laser technique was found to be fast and capable of attaining a diameter measurement accuracy of 0.1μm. © 1990 Chapman and Hall Ltd.
引用
收藏
页码:4694 / 4698
页数:5
相关论文
共 6 条
[1]  
Bohren C. F., 2008, ABSORPTION SCATTERIN
[2]  
FISCHBACH DB, 1988, COMMUNICATION
[3]  
GAGNAIRE P, 1987, J CHIM PHYS PCB, V84, P1407
[4]  
KOEDAM M, 1966, PHILIPS TECH RUND, V27, P1982
[5]   FIBER DIAMETER MEASUREMENT BY LASER DIFFRACTION [J].
PERRY, AJ ;
INEICHEN, B ;
ELIASSON, B .
JOURNAL OF MATERIALS SCIENCE, 1974, 9 (08) :1376-1378
[6]  
PERRY AJ, 1971, FIBRE SCI TECHNOL, V3, P317