COMPUTER-SIMULATION AS A TOOL FOR THE RAPID OPTIMIZATION OF THE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION OF A TRYPTIC DIGEST OF HUMAN GROWTH-HORMONE

被引:29
作者
CHLOUPEK, RC [1 ]
HANCOCK, WS [1 ]
SNYDER, LR [1 ]
机构
[1] LC RESOURCES INC,WALNUT CREEK,CA 94596
来源
JOURNAL OF CHROMATOGRAPHY | 1992年 / 594卷 / 1-2期
关键词
D O I
10.1016/0021-9673(92)80313-J
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Computer simulation was used to optimize the separation of a tryptic digest of recombinant human growth hormone using reversed-phase high-performance liquid chromatography in a gradient mode. DryLab G/plus software modelled the retention behavior of the complex tryptic digest mixture as a function of gradient conditions, based on data from two experimental gradient runs. The theoretical optimum separation conditions were rapidly obtained and reproduced experimentally. Resolution did not simply increase as gradient steepness was decreased, rather, an intermediate gradient time provided maximum sample resolution. The simulation results also indicate that the method is reasonably rugged, with little change in the separation expected for different high-performance liquid chromatography systems, and changes in the separation can be compensated by a change in the gradient steepness. Computer simulation can also be useful to quickly reoptimize conditions for a new column, if it fails to provide the same separation.
引用
收藏
页码:65 / 73
页数:9
相关论文
共 24 条
[1]  
BENNETT WF, 1989, ADV GROWTH HORMONE G, P29
[2]   SELECTION OF MOBILE PHASE PARAMETERS AND THEIR OPTIMIZATION IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
BILLIET, HAH ;
DEGALAN, L .
JOURNAL OF CHROMATOGRAPHY, 1989, 485 :27-50
[3]  
BURSTYN DG, 1991, BIOPHARM TECHNOL FEB, P22
[4]   DRYLAB COMPUTER-SIMULATION FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD DEVELOPMENT .2. GRADIENT ELUTION [J].
DOLAN, JW ;
LOMMEN, DC ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHY, 1989, 485 :91-112
[5]  
DOLAN JW, 1990, 14TH INT S COL LIQ C
[6]   THE ROLE OF QUALITY-CONTROL IN BIOTECHNOLOGY - AN ANALYTICAL PERSPECTIVE [J].
GARNICK, RL ;
SOLLI, NJ ;
PAPA, PA .
ANALYTICAL CHEMISTRY, 1988, 60 (23) :2546-2557
[7]   DESIGN OF OPTIMIZED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC GRADIENTS FOR THE SEPARATION OF EITHER SMALL OR LARGE MOLECULES .2. BACKGROUND AND THEORY [J].
GHRIST, BFD ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHY, 1988, 459 :25-41
[8]   DESIGN OF OPTIMIZED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC GRADIENTS FOR THE SEPARATION OF EITHER SMALL OR LARGE MOLECULES .3. AN OVERALL STRATEGY AND ITS APPLICATION TO SEVERAL EXAMPLES [J].
GHRIST, BFD ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHY, 1988, 459 :43-63
[9]   DESIGN OF OPTIMIZED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC GRADIENTS FOR THE SEPARATION OF EITHER SMALL OR LARGE MOLECULES .1. MINIMIZING ERRORS IN COMPUTER-SIMULATIONS [J].
GHRIST, BFD ;
COOPERMAN, BS ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHY, 1988, 459 :1-23
[10]  
GHRIST BFD, 1990, HPLC BIOL MACROMOLEC, pCH15