POSITIONING AND ENERGY RESPONSE OF PET BLOCK DETECTORS WITH DIFFERENT LIGHT SHARING SCHEMES

被引:35
作者
TORNAI, MP
GERMANO, G
HOFFMAN, EJ
机构
[1] CEDARS SINAI MED CTR, LOS ANGELES, CA 90048 USA
[2] UNIV CALIF LOS ANGELES, SCH MED, DEPT PHARMACOL, LOS ANGELES, CA 90048 USA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.322931
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Two state-of-the-art modular PET block detectors using either discrete or pseudo-discrete BGO crystals coupled to two dual PMTs, utilizing different light sharing schemes, were evaluated. Both detectors were approximately 30 mm thick, while each element of the GE 6 x 6 block detector was 4 mm x 8.4 mm, and each element of the CTI 8 x 7 block detector was 2.8 mm x 5.8 mm. In addition to measurements with gamma sources, the detectors were also irradiated with a Sr/Y-90 beta source to evaluate the performance of the block detectors without inter-element scatter of annihilation photons. The best and worst energy resolutions of individual elements at 511 keV were 21.8% and 44.2% for the GE detector, and 22.7% and 42.5% for the CTI detector. The peak to valley ratios in the detector identification ratio histograms were generally better than 3 to 1. Measurements with beta sources indicate that light sharing in both detectors is a large component of event mispositioning along with inter-detector scatter of the annihilation photons.
引用
收藏
页码:1458 / 1463
页数:6
相关论文
共 8 条
[1]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[2]   AN EVALUATION OF A TWO-DIMENSIONAL ARRAY DETECTOR FOR HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (04) :264-272
[3]   AN INVESTIGATION OF METHODS OF PILEUP REJECTION FOR 2-D ARRAY DETECTORS EMPLOYED IN HIGH-RESOLUTION PET [J].
GERMANO, G ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (02) :223-227
[4]  
MOSES WW, 1993, J NUCL MED, V34, pP101
[5]   AN IMPROVED MULTICRYSTAL 2-D BGO DETECTOR FOR PET [J].
ROGERS, JG ;
TAYLOR, AJ ;
RAHIMI, MF ;
NUTT, R ;
ANDREACO, M ;
WILLIAMS, CW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :1063-1068
[6]  
TORNAI MP, 1994, IN PRESS NUCL INST A
[7]  
UCHIDA H, 1992, Patent No. 5091650
[8]   A POSITRON CAMERA DETECTOR DESIGN WITH CROSS-COUPLED SCINTILLATORS AND QUADRANT SHARING PHOTOMULTIPLIERS [J].
WONG, WH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :962-966