摘要
Equal time spacing of RF pulses in the CPMG sequence imposes a constraint of equal signal read periods in spin-echo train imaging. GRASE imaging differs by using multiple read gradients in each pi-pi time interval, which are not constrained to be equal in number or duration. This additional degree of freedom is developed in dual contrast imaging. Closely spaced read periods are used for the PDW image to reduce T2 decay effects, while fewer low-bandwidth read periods in each of several pi-pi intervals are used to raise the signal-to-noise ratio and avoid signal averaging in the T2-weighted image.
引用
收藏
页码:461 / 464
页数:4
相关论文
共 16 条
[1]  
BORNERT P, 1993, 12TH P SMRM ANN M, P1235
[2]   THE LOSS OF SMALL OBJECTS IN VARIABLE TE IMAGING - IMPLICATIONS FOR FSE, RARE, AND EPI [J].
CONSTABLE, RT ;
GORE, JC .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) :9-24
[3]  
CROOKS LE, 1982, RADIOLOGY, V143, P143
[4]  
EDELSTEIN WA, 1980, Patent No. 2079463
[5]   GRASE (GRADIENT-ECHO AND SPIN-ECHO) MR IMAGING - A NEW FAST CLINICAL IMAGING TECHNIQUE [J].
FEINBERG, DA ;
OSHIO, K .
RADIOLOGY, 1991, 181 (02) :597-602
[6]   GRADIENT-ECHO SHIFTING IN FAST MRI TECHNIQUES (GRASE IMAGING) FOR CORRECTION OF FIELD INHOMOGENEITY ERRORS AND CHEMICAL-SHIFT [J].
FEINBERG, DA ;
OSHIO, K .
JOURNAL OF MAGNETIC RESONANCE, 1992, 97 (01) :177-183
[7]  
FEINBERG DA, 1991, Patent No. 5270654
[8]  
FRAM EK, 1991, 10TH P SMRM ANN M SA, P218
[9]  
HENNIG J, 1986, J COMPUT ASSIST TOMO, V10, P375
[10]  
KIEFER B, 1992, 11TH P SMRM ANN M BE, P4523