CALCULATING THE CALIBRATION COEFFICIENT FOR RADON MEASUREMENTS WITH THE BARE LR 115-II TRACK DETECTOR

被引:26
作者
NIKEZIC, D
MARKOVIC, P
UZAROV, DB
机构
[1] Kragujevac
来源
HEALTH PHYSICS | 1992年 / 62卷 / 03期
关键词
RN-222; CALIBRATION; DETECTOR; ETCHED-TRACK; MONTE-CARLO;
D O I
10.1097/00004032-199203000-00005
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Model calculations were performed to determine the calibration coefficient for radon measurements using a bare LR 115-II track detector. Calculations were based on an expression for stopping power for heavily charged particles, as well as on the Monte Carlo method. The sensitivity of the LR 115-II detector depends on equilibrium factor F, and it has been calculated for various values of F. For equilibrium factor F = 1, the calibration coefficient in this calculation was 89 m-1 (10.3 Bq d m-3 per tracks cm-2) for the removed layer, which was 5.5-mu-m thick.