OPTIMIZATION OF THE MASS SCANNING RATE FOR THE DETERMINATION OF LEAD ISOTOPE RATIOS USING AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER

被引:84
作者
FURUTA, N
机构
[1] Division of Environmental Chemistry, National Institute for Environmental Studies, Tsukuba, Ibaraki 305
关键词
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; LEAD ISOTOPE RATIO; NOISE AMPLITUDE SPECTRA; HIGH-SPEED MOTION PICTURES; DWELL TIME;
D O I
10.1039/ja9910600199
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High-speed motion pictures of a plasma and the noise-amplitude spectra of the emission signals enabled clarification that the inductively coupled plasma coupled with a mass spectrometer had a noise component at a frequency of 380 Hz under the experimental conditions used. In order to improve the precision of lead isotope ratio measurements the mass scanning rate was optimized. With an increase in the mass scanning rate, the precision of lead isotope ratio measurements was improved because the low frequency noises could be eliminated. However, the quadrupole mass analyser could not be operated at a speed that was sufficient to smooth out the audio frequency noise.
引用
收藏
页码:199 / 203
页数:5
相关论文
共 13 条
[1]   ACCURATE AND PRECISE DETERMINATION OF LEAD ISOTOPE RATIOS IN CLINICAL AND ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY [J].
CAMPBELL, MJ ;
DELVES, HT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (02) :235-236
[2]   NOISE POWER SPECTRAL CHARACTERISTICS OF AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
CRAIN, JS ;
HOUK, RS ;
ECKELS, DE .
ANALYTICAL CHEMISTRY, 1989, 61 (06) :606-612
[3]   SELECTION OF MODE FOR THE MEASUREMENT OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY AND ITS APPLICATION TO MILK POWDER ANALYSIS [J].
DEAN, JR ;
EBDON, L ;
MASSEY, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (04) :369-374
[4]   NOISE CHARACTERISTICS OF AN INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETER [J].
FURUTA, N ;
MONNIG, CA ;
YANG, PY ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (07) :649-656
[5]   SPATIALLY RESOLVED NOISE AMPLITUDE SPECTRA OF EMISSION SIGNALS FROM AN INDUCTIVELY COUPLED PLASMA [J].
FURUTA, N .
ANALYTICAL SCIENCES, 1990, 6 (05) :683-688
[6]   INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRIC DETERMINATION OF LEAD ISOTOPES [J].
HINNERS, TA ;
HEITHMAR, EM ;
SPITTLER, TM ;
HENSHAW, JM .
ANALYTICAL CHEMISTRY, 1987, 59 (22) :2658-2662
[7]   DETERMINATION OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY (ICP-MS) [J].
LONGERICH, HP ;
FRYER, BJ ;
STRONG, DF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (1-2) :39-48
[8]  
MUKAI H, 1990, BUNSEKI KAGAKU, V39, P177
[9]   ISOTOPIC RATIO MEASUREMENTS WITH AN INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETER [J].
RUSS, GP ;
BAZAN, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (1-2) :49-62
[10]   LEAD-206/207 ISOTOPE RATIOS IN THE ATMOSPHERE OF NORTH-AMERICA AS TRACERS OF UNITED-STATES AND CANADIAN EMISSIONS [J].
STURGES, WT ;
BARRIE, LA .
NATURE, 1987, 329 (6135) :144-146