PROGRAMMING OF CARDIAC-PACEMAKERS - WISH AND REALITY

被引:12
作者
IRNICH, W
KRAMER, E
MULLER, R
机构
[1] Inst. Med. Tech. Universitat, W-6300 Giessen
关键词
D O I
10.1055/s-2008-1063654
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the extent and quality of pacemaker programming, 340 programmable pacemakers were selected out of 1,204 impulse generators removed post-mortem. Their actual program was compared with the nominal parameters when first delivered. 177 (52.1%) of the 340 pacema kers still had the same parameters as on delivery. The multiply programmable impulse generators with telemetry were more frequently programmed (61.2%) than the simple programmable ones (43.0%) or the multiply programmable ones without telemetry (43.7%). Rate was the most frequently programmed parameter, followed by impulse amplitude, sensitivity and impulse duration. Refractory time and hysteresis were only rarely changed from the initial state. Apparently, programmability was not thought to be important enough. Programming without telemetry was seemingly too complicated. The reasons for this attitude should be investigated so that the discrepancy between wishes and reality can be overcome.
引用
收藏
页码:601 / 605
页数:5
相关论文
共 11 条
[1]  
BAROLD SS, 1981, BRIT HEART J, V45, P357
[2]   UNIVERSAL PROGRAMMING [J].
FURMAN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (02) :131-132
[3]   PROGRAMMING PACEMAKERS - A GAIN OR A GIMMICK [J].
IRNICH, W .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1985, 110 (40) :1527-1530
[4]   THE CHRONAXIE TIME AND ITS PRACTICAL IMPORTANCE [J].
IRNICH, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (03) :292-301
[5]  
IRNICH W, 1989, HERZSCHRITTMACHER, V9, P164
[6]  
IRNICH W, 1990, HERZSCHRITTMACHER 19, V10, P138
[7]   EXPERIENCE WITH TELEMETRY-SUPPORTED PACEMAKER CONTROL IN PATIENTS WITH VVI-PACEMAKERS [J].
KLEIN, HH ;
HANH, BK ;
HELLBERG, K ;
RUSCHEWSKI, W ;
DEVIVIE, ER ;
KREUZER, H .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1985, 110 (38) :1447-1451
[8]  
LASCAULT G, 1988, Presse Medicale, V17, P1117
[9]   CLINICAL-SIGNIFICANCE OF CARDIAC-PACEMAKER PROGRAMMING [J].
LUDERITZ, B .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1985, 110 (40) :1519-1520
[10]   IMPORTANCE AND LONG-TERM UTILITY OF A MULTIPARAMETER PROGRAMMABLE PULSE-GENERATOR [J].
OKEEFE, JH ;
HAYES, DL ;
HOLMES, DR ;
VLIETSTRA, RE .
MAYO CLINIC PROCEEDINGS, 1984, 59 (04) :239-242