POLARITY EFFECT FOR VARIOUS IONIZATION CHAMBERS WITH MULTIPLE IRRADIATION CONDITIONS IN ELECTRON-BEAMS

被引:23
作者
AGET, H [1 ]
ROSENWALD, JC [1 ]
机构
[1] INST CURIE,MED SECT,UNITE RADIOPHYS,F-75231 PARIS 05,FRANCE
关键词
ELECTRONS; POLARITY EFFECT; IONIZATION CHAMBERS;
D O I
10.1118/1.596694
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of reversing the voltage polarity applied to an ionization chamber has been investigated in electron beams for several types of chambers and several irradiation conditions. It has been found that differences in readings can be significant for cylindrical chambers (about 10%) as well as for plane parallel chambers (20%). The effect is larger for large field sizes than small ones. It generally includes an appreciable stem and cable effect. Differences in readings with both polarities are related to the energy distribution of the electron beam and are greater for lower electron energies than higher. Polarity effect and charge deposit within the chamber wall material appear to be closely connected. This charge deposit, expressed as a proportion of the total collected charge, can be directly derived from double polarities measurements. Careful investigation of the effect should be made to avoid significant error (over 5%) in the determination of the absorbed dose.
引用
收藏
页码:67 / 72
页数:6
相关论文
共 9 条
[1]   THE POLARITY EFFECT FOR COMMERCIALLY AVAILABLE PLANE-PARALLEL IONIZATION CHAMBERS [J].
GERBI, BJ ;
KHAN, FM .
MEDICAL PHYSICS, 1987, 14 (02) :210-215
[2]   CURRENTS INDUCED IN THE DIELECTRICS OF IONIZATION CHAMBERS THROUGH THE ACTION OF HIGH-ENERGY RADIATION [J].
JOHNS, HE ;
ASPIN, N ;
BAKER, RG .
RADIATION RESEARCH, 1958, 9 (06) :573-588
[3]  
MARINELLO G, 1981, J EUR RADIOTHER, V2, P37
[4]   CALIBRATION AND USE OF PLANE-PARALLEL IONIZATION CHAMBERS FOR THE DETERMINATION OF ABSORBED DOSE IN ELECTRON-BEAMS [J].
MATTSSON, LO ;
JOHANSSON, KA ;
SVENSSON, H .
ACTA RADIOLOGICA ONCOLOGY, 1981, 20 (06) :385-399
[5]   EFFECT OF CHAMBER VOLTAGE ON ELECTRON BUILD-UP MEASUREMENTS [J].
RICHARDSON, JE .
RADIOLOGY, 1954, 62 (04) :584-588
[6]   INVESTIGATION OF CABLES FOR IONIZATION CHAMBERS [J].
SPOKAS, JJ ;
MEEKER, RD .
MEDICAL PHYSICS, 1980, 7 (02) :135-140
[7]  
VANDYK J, 1972, PHYS MED BIOL, V17, P52
[8]   CHARGE DEPOSITION FROM HIGH-ENERGY ELECTRON-BEAMS [J].
VANDYK, J .
RADIATION RESEARCH, 1972, 50 (01) :20-+
[9]  
1986, ICRU35 REP