ACCURACY AND ANALYSIS OF AMBULATORY BLOOD-PRESSURE MONITORING DATA

被引:7
作者
WHITE, WB
机构
[1] Section of Hypertension and Vascular Diseases, University of Connecticut School of Medicine, Farmington, Connecticut
关键词
AMBULATORY MONITORING; AREA UNDER CURVE (AUC); BLOOD PRESSURE LOAD; LEFT VENTRICULAR HYPERTROPHY; BLOOD PRESSURE PROFILE;
D O I
10.1002/clc.4960151405
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Two devices used to record blood pressure, the ambulatory blood pressure monitoring recorder and standard stethoscope and mercury column, were tested for accuracy against the direct intra-arterial blood pressure of patients at rest and during exercise. Recorders were found to be as accurate as mercury column measurement in patients at rest. A number of assessment techniques of ambulatory data are reviewed, including: calculation of mean or median pressures, assessment of blood pressure load, and integration of the area under the blood pressure curve over time. These have been applied during the daytime and nighttime hours. Blood pressure load and area under the blood pressure curve, using different threshold criteria for nighttime and daytime, are recommended because of their potentially closer relation to target-organ disease of hypertension than are office blood pressure readings.
引用
收藏
页码:10 / 13
页数:4
相关论文
共 6 条
[1]   WHAT IS THE VALUE OF HOME BLOOD-PRESSURE MEASUREMENT IN PATIENTS WITH MILD HYPERTENSION [J].
KLEINERT, HD ;
HARSHFIELD, GA ;
PICKERING, TG ;
DEVEREUX, RB ;
SULLIVAN, PA ;
MARION, RM ;
MALLORY, WK ;
LARAGH, JH .
HYPERTENSION, 1984, 6 (04) :574-578
[2]   HOW COMMON IS WHITE COAT HYPERTENSION [J].
PICKERING, TG ;
JAMES, GD ;
BODDIE, C ;
HARSHFIELD, GA ;
BLANK, S ;
LARAGH, JH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 259 (02) :225-228
[3]   USEFULNESS OF AMBULATORY MONITORING OF BLOOD-PRESSURE IN ASSESSING ANTIHYPERTENSIVE THERAPY [J].
WHITE, WB ;
MORGANROTH, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (01) :94-98
[4]   ASSESSMENT OF THE DAILY BLOOD-PRESSURE LOAD AS A DETERMINANT OF CARDIAC-FUNCTION IN PATIENTS WITH MILD-TO-MODERATE HYPERTENSION [J].
WHITE, WB ;
DEY, HM ;
SCHULMAN, P .
AMERICAN HEART JOURNAL, 1989, 118 (04) :782-795
[5]   ASSESSMENT OF 4 AMBULATORY BLOOD-PRESSURE MONITORS AND MEASUREMENTS BY CLINICIANS VERSUS INTRAARTERIAL BLOOD-PRESSURE AT REST AND DURING EXERCISE [J].
WHITE, WB ;
LUNDJOHANSEN, P ;
OMVIK, P .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 65 (01) :60-66
[6]  
WHITE WB, 1991, J HYPERTENS, V9, pS27