PRECISION AND ACCURACY OF ANALYSIS OF AIR-FILLED ALBUMIN MICROSPHERES USING COULTER-MULTISIZER-MARK-II

被引:16
作者
SONTUM, PC
CHRISTIANSEN, C
机构
[1] Research and Development Section, Nycomed Imaging AS, N-0401 Oslo
关键词
COULTER MULTISIZER; ELECTRICAL SENSING ZONE TECHNIQUE; PRECISION; ACCURACY; AIR-FILLED MICROSPHERES; ALBUMIN;
D O I
10.1016/0731-7085(94)00069-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The suitability of the electrical sensing zone technique for routine analysis of air-filled albumin microspheres has been thoroughly investigated using three Coulter Multisizer Mark II instruments. The precision of the method, expressed as repeatability relative standard deviation (RSD), was found to be 1-2% for number distribution parameters and 3-4% for Volume distribution parameters. Significant instrument-to-instrument variation as high as 11% was, however, also observed. Accuracy was evaluated from analyses of commercially available latex standards and from comparison of results from Coulter analysis with results from the following alternative techniques: light diffraction, optical microscopy and gravimetry. Accuracy, expressed as the difference from either certified values or values obtained with the alternative techniques, was found to be 100-106%.