MULTILEAF COLLIMATION VERSUS ALLOY BLOCKS - ANALYSIS OF GEOMETRIC ACCURACY

被引:33
作者
LOSASSO, T
KUTCHER, GJ
机构
[1] Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY 10021
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1995年 / 32卷 / 02期
关键词
MULTILEAF COLLIMATOR; CUSTOM BLOCKS; GEOMETRIC ACCURACY;
D O I
10.1016/0360-3016(94)00455-T
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To compare the misalignment error due to the fabrication of custom lead alloy blocks with the displacement error introduced by the finite resolution of a multileaf collimator relative to the prescribed smooth apertures. Methods and Materials: Treatment field apertures for randomly selected patients for four clinical sites were obtained at various stages of the block fabrication process. These apertures and the corresponding multileaf collimator (MLC) apertures for each field were superimposed with the smooth apertures prescribed by physicians. The deviations from the prescribed apertures were measured at 10 degrees intervals. Comparisons of the magnitude and frequency of errors from block fabrication with those from the geometric displacements introduced by the finite leaf width of the multileaf collimator were made. Results: The degree of conformity of the multileaf collimator is treatment-site dependent as, in general, are the shapes of fields. For three of the four sites examined, the multileaf collimator apertures track the prescribed apertures at least as accurately as custom blocking when the block design, construction, mounting,and alignment on the treatment machine are considered. Conclusions: The geometric conformality of multileaf collimation is comparable to, and in some cases superior to, that of custom blocks.
引用
收藏
页码:499 / 506
页数:8
相关论文
共 12 条
[1]   CLINICAL DOSIMETRY FOR IMPLEMENTATION OF A MULTILEAF COLLIMATOR [J].
BOYER, AL ;
OCHRAN, TG ;
NYERICK, CE ;
WALDRON, TJ ;
HUNTZINGER, CJ .
MEDICAL PHYSICS, 1992, 19 (05) :1255-1261
[2]   OPTIMIZATION OF RADIATION-THERAPY AND THE DEVELOPMENT OF MULTILEAF COLLIMATION [J].
BRAHME, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :373-375
[3]  
BRAHME A, 1988, STRAHLENTHER ONKOL, V164, P343
[4]  
CHUI CS, 1993, MED PHYS, V20, P885
[5]   EVALUATION OF MULTILEAF COLLIMATOR DESIGN FOR A PHOTON-BEAM [J].
GALVIN, JM ;
SMITH, AR ;
MOELLER, RD ;
GOODMAN, RL ;
POWLIS, WD ;
RUBENSTEIN, J ;
SOLIN, LJ ;
MICHAEL, B ;
NEEDHAM, M ;
HUNTZINGER, CJ ;
KLIGERMAN, MM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 23 (04) :789-801
[6]   COMBINING MULTILEAF FIELDS TO MODULATE FLUENCE DISTRIBUTIONS [J].
GALVIN, JM ;
CHEN, XG ;
SMITH, RM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (03) :697-705
[7]  
Leibel, 1992, Semin Radiat Oncol, V2, P274, DOI 10.1016/1053-4296(92)90025-G
[8]   THE USE OF A MULTI-LEAF COLLIMATOR FOR CONFORMAL RADIOTHERAPY OF CARCINOMAS OF THE PROSTATE AND NASOPHARYNX [J].
LOSASSO, T ;
CHUI, CS ;
KUTCHER, GJ ;
LEIBEL, SA ;
FUKS, Z ;
LING, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :161-170
[9]   A MODEL FOR COMPUTER-CONTROLLED DELIVERY OF 3-D CONFORMAL TREATMENTS [J].
MAGERAS, GS ;
PODMANICZKY, KC ;
MOHAN, R .
MEDICAL PHYSICS, 1992, 19 (04) :945-953
[10]   INITIAL CLINICAL-EXPERIENCE WITH COMPUTER-CONTROLLED CONFORMAL RADIOTHERAPY USING THE MM50 MICROTRON [J].
MAGERAS, GS ;
FUKS, Z ;
OBRIEN, J ;
BREWSTER, LJ ;
BURMAN, CM ;
CHUI, CS ;
LEIBEL, SA ;
LING, CC ;
MASTERSON, ME ;
MOHAN, R ;
KUTCHER, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 :207-208