STUDY ON THE USE OF TRANSMISSION SCANS FOR WHOLE-BODY PET ATTENUATION CORRECTION

被引:5
作者
CHATZIIOANNOU, A
DAHLBOM, M
HOH, CJ
机构
[1] Division of Nuclear Medicine, Department of Molecular and Medical Pharmacology UCLA School of Medicine, Los Angeles, CA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.322946
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Transmission scans for attenuation correction of whole body PET studies are not acquired routinely, since noise considerations impose acquisition time constraints that make conventional scanning techniques infeasible. The aim of this work is to optimize data acquisition and processing parameters and improve the SNR of whole body transmission scans, in order to achieve within a reasonable time frame, an attenuation correction of the same quality as in cardiac PET. Methods to improve the scanner sensitivity by using additional coincidence planes, as well as smoothing methods for the randoms and the transmission data were investigated. Phantom and patient studies show that transmission scans are feasible for whole body PET studies, with equivalent noise introduced in the chest area as in a typical cardiac PET attenuation correction, at fractions of the time required by conventional transmission scan protocols.
引用
收藏
页码:1545 / 1550
页数:6
相关论文
共 14 条
[1]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .7. A TECHNIQUE TO REDUCE NOISE IN ACCIDENTAL COINCIDENCE MEASUREMENTS AND COINCIDENCE EFFICIENCY CALIBRATION [J].
CASEY, ME ;
HOFFMAN, EJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1986, 10 (05) :845-850
[2]  
DAHLBOM M, 1992, J NUCL MED, V33, P1191
[3]   PROBLEMS IN SIGNAL-TO-NOISE RATIO FOR ATTENUATION CORRECTION IN HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :288-293
[4]   METHODS FOR IMPROVING IMAGE QUALITY IN WHOLE-BODY PET SCANNING [J].
DAHLBOM, M ;
YU, DC ;
CHERRY, SR ;
CHATZIIOANNOU, A ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :1079-1083
[5]   CHARACTERIZATION OF A WHOLE-BODY IMAGING TECHNIQUE FOR PET [J].
GUERRERO, TM ;
HOFFMAN, EJ ;
DAHLBOM, M ;
CUTLER, PD ;
HAWKINS, RA ;
PHELPS, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :676-680
[6]  
HAWKINS RA, 1991, J NUCL MED, V32, P1555
[7]   CANCER-DETECTION WITH WHOLE-BODY PET USING 2-[F-18]FLUORO-2-DEOXY-D-GLUCOSE [J].
HOH, CK ;
HAWKINS, RA ;
GLASPY, JA ;
DAHLBOM, M ;
TSE, NY ;
HOFFMAN, EJ ;
SCHIEPERS, C ;
CHOI, Y ;
REGE, S ;
NITZSCHE, E ;
MADDAHI, J ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1993, 17 (04) :582-589
[8]  
HUANG SC, 1979, J COMPUT ASSIST TOMO, V3, P804
[9]  
HUANG SC, 1981, J NUCL MED, V22, P627
[10]  
MEIKLE SR, 1993, J NUCL MED, V34, P143