NEW FIBER-OPTIC-BASED INSTRUMENTATION FOR THE MEASUREMENT OF LOW-TEMPERATURE PHOSPHORESCENCE INTENSITIES AND LIFETIMES

被引:9
作者
CARROLL, MK [1 ]
HIEFTJE, GM [1 ]
机构
[1] INDIANA UNIV,DEPT CHEM,BLOOMINGTON,IN 47405
关键词
PHOSPHORESCENCE; PHOSPHORESCENCE LIFETIME; TIME-RESOLVED SPECTROSCOPY; LUMINESCENCE; INSTRUMENTATION; FIBER OPTICS; ACETOPHENONE; BENZOPHENONE; PARA-NITROPHENOL;
D O I
10.1366/0003702924444326
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
New fiber-optic instrumentation for phosphorescence-lifetime determinations and for steady-state phosphorescence measurements is shown to perform comparably to conventional instrumentation. The importance of optimal glass formation in the fiber-optic sample cell is considered in some detail. Detection limits for acetophenone, benzophenone, and p-nitrophenol are reported, and phosphorescence lifetimes are measured for several compounds, with values ranging from 2.5 ms to 4.7 s.
引用
收藏
页码:126 / 130
页数:5
相关论文
共 18 条