SUITABILITY OF POLYSTYRENE FOR VOLTAMMETRIC CELLS - A DIFFERENTIAL-PULSE ANODIC-STRIPPING VOLTAMMETRIC STUDY

被引:14
作者
DIAZCRUZ, S [1 ]
DIAZCRUZ, JM [1 ]
ESTEBAN, M [1 ]
机构
[1] UNIV BARCELONA,DEPT ANALYT CHEM,E-08028 BARCELONA,SPAIN
关键词
D O I
10.1021/ac00081a030
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The convenience of using polystyrene as material for voltammetric cells in the determination of Zn, Cd, Pb, and Cu was studied by differential pulse anodic stripping voltammetry. Cells of glass, PTFE Teflon, colorless polystyrene, and white polystyrene were compared. Polystyrene seems to be more suitable than PTFE Teflon or glass, especially at pH > 6, in 0.01 mol L(-1) KNO3 medium. In 0.5 mol L(-1) NaCl medium, differences between PTFE Teflon and polystyrene are not so important. However, the, use of polystyrene cells seems especially appropriate for measurements of Pb(II) and Cu(II).
引用
收藏
页码:1548 / 1551
页数:4
相关论文
共 15 条
[1]  
Baes CF., 1976, HYDROLYSIS CATIONS
[2]  
Batley GE, 1989, TRACE ELEMENT SPECIA, P1
[3]   STRIPPING VOLTAMMETRY OF METAL-COMPLEXES - INTERFERENCES FROM ADSORPTION ONTO CELL COMPONENTS [J].
DIAZCRUZ, JM ;
ESTEBAN, M ;
VANDENHOOP, MAGT ;
VANLEEUWEN, HP .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1769-1776
[4]   STYRENE MONOMER MIGRATION AS MONITORED BY PURGE AND TRAP GAS-CHROMATOGRAPHY AND SENSORY ANALYSIS FOR POLYSTYRENE CONTAINERS [J].
DURST, GL ;
LAPERLE, EA .
JOURNAL OF FOOD SCIENCE, 1990, 55 (02) :522-524
[5]  
FISHBEIN L, 1984, ADDITIVES SYNTHETIC, P19
[6]   THE LOSS OF CADMIUM AND ZINC FROM SEA-WATER DURING ACCUMULATION EXPERIMENTS - ITS IMPLICATION ON TOXICITY THRESHOLD CONCENTRATIONS [J].
HENNIG, HFKO ;
GREENWOOD, PJ .
MARINE POLLUTION BULLETIN, 1981, 12 (02) :47-50
[7]  
ISO S, 1991, KOGYO YOSUI, V396, P33
[8]   CLEANING METHODS FOR POLYTHENE CONTAINERS PRIOR TO THE DETERMINATION OF TRACE-METALS IN FRESH-WATER SAMPLES [J].
LAXEN, DPH ;
HARRISON, RM .
ANALYTICAL CHEMISTRY, 1981, 53 (02) :345-350
[9]   THE EFFECTS OF STORAGE TIMES, TEMPERATURES AND CONTAINER TYPES ON THE ACCURACY OF ATOMIC-ABSORPTION DETERMINATIONS OF CD, CU, HG, PB AND ZN IN WHOLE HEPARINIZED BLOOD [J].
MERANGER, JC ;
HOLLEBONE, BR ;
BLANCHETTE, GA .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1981, 5 (01) :33-41
[10]   SELECTION AND CLEANING OF PLASTIC CONTAINERS FOR STORAGE OF TRACE-ELEMENT SAMPLES [J].
MOODY, JR ;
LINDSTROM, RM .
ANALYTICAL CHEMISTRY, 1977, 49 (14) :2264-2267