AN ASSESSMENT OF THE EFFECT OF HEMATOCRIT ON THE HEMOCUE BLOOD-GLUCOSE ANALYZER

被引:31
作者
WIENER, K
机构
[1] Department of Clinical Biochemistry, North Manchester General Hospital
关键词
NEAR-PATIENT TESTING; ACCURACY; PRECISION; GLUCOSE STRIP-TESTS;
D O I
10.1177/000456329303000116
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The HemoCue blood glucose analyser was investigated with particular respect to its susceptibility to influence by the haematocrit of the blood specimen. The red cells and plasma of fresh blood samples were separated and then recombined in proportions to give a range of haematocrit values and the resulting specimens assayed for glucose. The HemoCue exhibited no appreciable influence of haematocrit. The instrument was easy to use, stable and had good precision (CV 1.8%). However, results tended to be a little higher than those produced by a conventional laboratory analyser on whole blood and it is understood that the factory calibration of newer HemoCue instruments has been modified to correct for this.
引用
收藏
页码:90 / 93
页数:4
相关论文
共 6 条
[1]  
BANAUCH D, 1975, Z KLIN CHEM KLIN BIO, V13, P101
[2]   ARE BLOOD-GLUCOSE REAGENT STRIPS RELIABLE IN RENAL-FAILURE [J].
CLARK, JDA ;
GOLDBERG, L ;
JONES, K ;
HARTOG, M .
DIABETIC MEDICINE, 1991, 8 (02) :168-171
[3]   REFLECTANCE METER TYPE AND MANAGEMENT OF GESTATIONAL DIABETES [J].
HARMELIN, D ;
MOLYNEAUX, L ;
FOWLER, P ;
WILLEY, K ;
WAIBUTA, K ;
LAU, R ;
ROBIN, H ;
HOSKINS, P ;
FORREST, J ;
ELLWOOD, D ;
YUE, DK ;
TURTLE, JR .
LANCET, 1990, 335 (8691) :735-736
[4]  
LABIB M, 1990, LANCET, V335, P973, DOI 10.1016/0140-6736(90)91038-C
[5]  
WIENER K, 1990, LANCET, V335, P974
[6]   THE EFFECT OF HEMATOCRIT ON REAGENT STRIP TESTS FOR GLUCOSE [J].
WIENER, K .
DIABETIC MEDICINE, 1991, 8 (02) :172-175