EVALUATION OF PERFORMANCE-CHARACTERISTICS OF DISPOSABLE BAG-VALVE RESUSCITATORS

被引:7
作者
KISSOON, N [1 ]
NYKANEN, D [1 ]
TIFFIN, N [1 ]
FREWEN, T [1 ]
BRASHER, P [1 ]
机构
[1] CHILDRENS HOSP WESTERN ONTARIO,DEPT PAEDIAT,LONDON,ONTARIO,CANADA
关键词
DISPOSABLE EQUIPMENT; PERFORMANCE; PHYSICAL CHARACTERISTICS; RESUSCITATION; TIDAL VOLUME; LABORATORIES; MULTIVARIATE ANALYSIS;
D O I
10.1097/00003246-199101000-00022
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the performance characteristics of disposable bag-valve resuscitators. Design: Single and multitrial tests under simulated clinical conditions. Results of multitrial tests were analyzed by multivariate analysis of variance. Setting: A pulmonary laboratory in a university hospital. Methods: We evaluated eight DRs to determine: a) physical characteristics, including the potential for misassembly; b) F(IO)2 of 1.0 and flow rates of 10 L; c) tidal volume (VT) (male and female operators) using one and two hands; and d) cycling rates at 5-degrees, 20-degrees, and 48-degrees-C. Results: All resuscitators met the minimum requirements of the Canadian Standards Association and the American Society for Testing and Materials. However, the CPR and LSP could be assembled incorrectly for use. Three units (B Sav, Pulm, and MPR) delivered significantly < 80% oxygen (overall p < .05). Significant increases were seen in VT using two hands vs. one hand (1176 vs. 960 mL, p < .001) and males vs. females using two hands (1284 vs. 1102 mL, p < .02). One unit (B Sav) could only be used with two hands. The maximum cycling rates for all units increased significantly (p < .05) with an increase in temperature from -5-degrees to 20-degrees to 40-degrees-C. However, two units (Bag E and MPR) failed to deliver 20 breaths/min at -5-degrees-C, while the B Sav and Pulm valves became dysfunctional at 48-degrees-C. Conclusions: While these DRs meet the minimum standards, they should be chosen carefully and evaluated before application in the clinical environment.
引用
收藏
页码:102 / 107
页数:6
相关论文
共 16 条
[1]  
BAKER PO, 1983, RESPIR CARE, V28, P1563
[2]   IDIOPATHIC DESTRUCTIVE ARTHRITIS OF THE SHOULDER [J].
CAMPION, GV ;
MCCRAE, F ;
ALWAN, W ;
WATT, I ;
BRADFIELD, J ;
DIEPPE, PA .
SEMINARS IN ARTHRITIS AND RHEUMATISM, 1988, 17 (04) :232-245
[3]   AN EVALUATION OF EMERGENCY MEDICAL TECHNICIANS ABILITY TO USE MANUAL VENTILATION DEVICES [J].
ELLING, R ;
POLITIS, J .
ANNALS OF EMERGENCY MEDICINE, 1983, 12 (12) :765-768
[4]  
FINER NN, 1986, PEDIATRICS, V77, P417
[5]  
HESS D, 1987, Respiratory Care, V32, P1025
[6]  
HESS D, 1986, Respiratory Care, V31, P774
[7]   BAG-VALVE-MASK VENTILATION - 2 RESCUERS ARE BETTER THAN ONE - PRELIMINARY-REPORT [J].
JESUDIAN, MCS ;
HARRISON, RR ;
KEENAN, RL ;
MAULL, KI .
CRITICAL CARE MEDICINE, 1985, 13 (02) :122-123
[8]   EVALUATION OF MASK-BAG VENTILATION IN RESUSCITATION OF INFANTS [J].
KANTER, RK .
AMERICAN JOURNAL OF DISEASES OF CHILDREN, 1987, 141 (07) :761-763
[9]  
LANGREHR EA, 1981, HEART LUNG, V10, P1028
[10]  
Law GD, 1982, RESP CARE, V27, P1236