ABERRATIONS IN CONFOCAL FLUORESCENCE MICROSCOPY INDUCED BY MISMATCHES IN REFRACTIVE-INDEX

被引:531
作者
HELL, S
REINER, G
CREMER, C
STELZER, EHK
机构
[1] EMBL, PHYS INSTRUMENTAT PROGRAMM, LIGHT MICROSCOPY GRP, MEYERHOFSTR 1, W-6900 HEIDELBERG, GERMANY
[2] UNIV HEIDELBERG, INST APPL PHYS, W-6900 HEIDELBERG, GERMANY
关键词
CONFOCAL FLUORESCENCE MICROSCOPY; REFRACTIVE-INDEX MISMATCH; VECTORIAL THEORY; ABERRATED PSF; FOCAL SHIFT; INTENSITY DROP; RESOLUTION; Z-RESPONSE;
D O I
10.1111/j.1365-2818.1993.tb03315.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
The effect of refractive-index mismatch, as encountered in the observation of biological specimens, on the image acquisition process in confocal fluorescence microscopy is investigated theoretically. The analysis takes the vectorial properties of light into account and is valid for high numerical apertures. Quantitative predictions on the decrease of resolution, intensity drop and shift of focus are given for practical situations. When observing with a numerical aperture of 1.3 (oil immersion) and an excitation wavelength of 514 nm the centre of the focus shifts 1.7 mum per 10 mum of axial displacement in an aqueous medium, thus yielding an image that is scaled by a factor of 1.2 in the axial direction. Furthermore, it can be expected that for a fluorescent plane 20 mum deep inside an aqueous medium the peak intensity is 40% less than for a plane which is 10 mum deep. In addition, the axial resolution is decreased by a factor of 1.4. The theory was experimentally verified for test samples with different refractive indices.