A LOCAL COINCIDENCE TRIGGERING SYSTEM FOR PET TOMOGRAPHS COMPOSED OF LARGE-AREA POSITION-SENSITIVE DETECTORS

被引:14
作者
MANKOFF, DA
MUEHLLEHNER, G
MILES, GE
机构
[1] UGM Medical Systems, Philadelphia, PA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.106706
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A local triggering system composed of small overlapping triggering channels has been designed for use in PET tomographs composed of continuous position-sensitive detectors. This triggering system functionally divides the continuous detectors into smaller sub-detectors for the purposes of event detection and coincidence timing. Such a system is necessary in tomographs with a continuous detector ring, such as a cylindrical NaI(T1) detector, where it is necessary to detect two coincident events in the same detector. A preliminary version of the local triggering system has been designed, and the effect of triggering channel size and overlap has been tested on a single large-area NaI(T1) detector. Based on these tests it is concluded that such a coincidence triggering system is feasible and that the local triggering system may reduce coincidence triggering system deadtime significantly. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.