ANALYSIS OF CLENBUTEROL IN HUMAN PLASMA USING LIQUID-CHROMATOGRAPHY ATMOSPHERIC-PRESSURE CHEMICAL-IONIZATION MASS-SPECTROMETRY

被引:37
作者
DOERGE, DR [1 ]
BAJIC, S [1 ]
LOWES, S [1 ]
机构
[1] VG BIOTECH LTD,FISONS INSTRUMENTS,ALTRINCHAM WA14 5RZ,CHESHIRE,ENGLAND
关键词
D O I
10.1002/rcm.1290070612
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Clenbuterol is a beta-agonist drug used illegally as a growth stimulant in meat-producing animals and human athletes. The analysis of clenbuterol in spiked human plasma was performed using on-line liquid chromatography/atmospheric-pressure chemical-ionization mass spectrometry (LC/APCI-MS) using a conventional bore LC column (flow rate = 1.0 mL/min). At low sampling cone voltages, the mass spectrum was predominantly the [M + H]+ ion but diagnostic fragment ions were formed upon incremental increases in sampling cone voltage. The detection limit (signal-to-noise ratio of 3) using selected-ion monitoring of the [M + H]+ ion was 0.1 ng on-column (10 ppb). This is a 50-fold better sensitivity of detection than that previously reported for an on-line thermospray LC/MS method. The extraction procedures were not optimized for maximum sensitivity and the lack of interferences suggests that much lower detection limits for clenbuterol in plasma are attainable.
引用
收藏
页码:462 / 464
页数:3