A HIGH-RESOLUTION XRII-BASED QUANTITATIVE VOLUME CT SCANNER

被引:68
作者
HOLDSWORTH, DW
DRANGOVA, M
FENSTER, A
机构
[1] UNIV WESTERN ONTARIO,DEPT DIAGNOST RADIOL,LONDON N6A 5B8,ON,CANADA
[2] UNIV WESTERN ONTARIO,DEPT MED BIOPHYS,LONDON N6A 5B8,ON,CANADA
关键词
QUANTITATIVE COMPUTED TOMOGRAPHY; IMAGE INTENSIFIER; 3-DIMENSIONAL RECONSTRUCTION;
D O I
10.1118/1.597038
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A laboratory volume CT scanner has been developed, with high spatial resolution in all three dimensions, which can be used for quantitative analysis of excised tissue samples in vitro. The system incorporates an x-ray image intensifier, optically coupled to a time-delay integration (TDI) CCD to obtain low-noise and low-scatter projections of the sample volume. A water bath surrounds the sample to equalize the exposure to the image intensifier, thereby reducing the dynamic range of the input signal. The scanner operates in two modes, producing either a single, transverse image through the sample or a three-dimensional image of the sample volume. Spatial resolution is adjustable over the range of 1.2 to 2.8 mm-1. System response is linear over the range -1000 to 3 500 Houndsfield units (HU), with an average precision of +/-80 HU. The precision of geometric measurements in the transverse plane allows circumference measurements to within +/-0.1 mm. Finally, applications of this technique of nondestructive analysis in biomedical research are discussed.
引用
收藏
页码:449 / 462
页数:14
相关论文
共 56 条
  • [1] CAPABILITY OF FLUOROSCOPIC SYSTEMS FOR PRODUCTION OF COMPUTERIZED AXIAL TOMOGRAMS
    BAILY, NA
    KELLER, RA
    JAKOWATZ, CV
    KAK, AC
    [J]. INVESTIGATIVE RADIOLOGY, 1976, 11 (05) : 434 - 439
  • [2] BARBE DF, 1976, SOLID STATE IMAGING, P659
  • [3] BARNES GT, 1979, P SOC PHOTO-OPT INS, V173, P225
  • [4] ANALYSIS AND CORRECTION OF IMPERFECTIONS IN THE IMAGE INTENSIFIER-TV-DIGITIZER IMAGING CHAIN
    BOONE, JM
    SEIBERT, JA
    BARRETT, WA
    BLOOD, EA
    [J]. MEDICAL PHYSICS, 1991, 18 (02) : 236 - 242
  • [5] BOONE JM, 1990, RADIOLOGY, V177, P136
  • [6] BROOKS R A, 1976, Medical Physics (Woodbury), V3, P237, DOI 10.1118/1.594240
  • [7] ALIASING - SOURCE OF STREAKS IN COMPUTED TOMOGRAMS
    BROOKS, RA
    GLOVER, GH
    TALBERT, AJ
    EISNER, RL
    DIBIANCA, FA
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (04) : 511 - 518
  • [8] COMPUTED TOMOGRAPHIC EVALUATION OF CERVICAL CAROTID PLAQUE COMPLICATIONS
    CULEBRAS, A
    LEESON, MD
    CACAYORIN, ED
    HODGE, CJ
    ILIYA, AR
    [J]. STROKE, 1985, 16 (03) : 425 - 431
  • [9] MECHANICAL-PROPERTIES OF ARTERIES
    DOBRIN, PB
    [J]. PHYSIOLOGICAL REVIEWS, 1978, 58 (02) : 397 - 460
  • [10] DRANGOVA M, 1992, RADIOLOGY, V185, P201