DETERMINATION OF PLASMA TESTOSTERONE BY MASS FRAGMENTOGRAPHY USING TESTOSTERONE-19-D3 AS AN INTERNAL STANDARD - COMPARISON WITH RADIOIMMUNOASSAY

被引:37
作者
BABA, S
SHINOHARA, Y
KASUYA, Y
机构
[1] Tokyo College of Pharmacy, Hachioji, Tokyo, 192-03
来源
JOURNAL OF CHROMATOGRAPHY | 1979年 / 162卷 / 04期
关键词
D O I
10.1016/S0378-4347(00)81832-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Analytical procedures for the measurement of testosterone by mass fragmentography (MF) using trideuterated testosterone (testosterone-19,19,19-d3) are described. For the calculation of plasma testosterone, peak height ratios were measured by MF performed on the molecular ions of the TFA derivative of testosterone (m/e 480) and testosterone-19,19,19-d3 (m/e 483). The sensitivity of the method was judged from the lower limit of detection of the mass spectrometer which was at 10 pg. For the measurement of the precision, the inter- and intra-assay coefficients of variation (C.V.) were calculated by using a pooled plasma sample; they were 3.15% and 1.79%, respectively. The specificity was investigated by the use of 5α-dihydrotestosterone and the MF method was found to afford a highly selective technique. These results obtained by MF have been compared with the results obtained by a radioimmunoassay method. © 1979.
引用
收藏
页码:529 / 537