RADIOIMMUNOASSAY OF HUMAN EOSINOPHIL CATIONIC PROTEIN (ECP) BY AN IMPROVED METHOD - ESTABLISHMENT OF NORMAL LEVELS IN SERUM AND TURNOVER INVIVO

被引:189
作者
PETERSON, CGB
ENANDER, I
NYSTRAND, J
ANDERSON, AS
NILSSON, L
VENGE, P
机构
[1] PHARM DIAGNOST AB,DEPT RES,INFLAMMAT MARKERS,UPPSALA,SWEDEN
[2] UNIV HOSP UPPSALA,DEPT THORAC SURG,S-75185 UPPSALA,SWEDEN
关键词
D O I
10.1111/j.1365-2222.1991.tb00847.x
中图分类号
R392 [医学免疫学];
学科分类号
100102 ;
摘要
A radiommunoassay was developed allowing measurement of the cytotoxic cationic ECP. The assay, which has a total incubation time of 3.5 hr, is a double antibody assay with radiolabelled ECP, covering the concentration range of 2-200-mu-g/l. Performance data show a detection limit of < 2-mu-g/l and a cross-reactivity with eosinophil protein X (EPX/EDN) of < 0.06%. The coefficient of variation (%) within the measuring range was, within assay 4.8-10.4, and total 6.6-12.0. The assay is useful for measurement in various body fluids including serum, nasal secretions and bronchoalveolar lavage fluid, and dilution of samples prior to analysis was generally not required. Sera from 100 apparently healthy individuals revealed a geometric mean of 6.0-mu-g ECP/l and a range (95%) of 2.3-15.9-mu-g/l. The elimination rate of ECP, t1/2, in vivo was estimated to be 65 min when ECP was measured in serum. Comparisons between this assay and a method previously described showed that the new method is superior with regard to precision and assay procedure.