CHARACTERIZATION OF BIOLOGICAL MACROMOLECULES BY COMBINED MASS MAPPING AND ELECTRON ENERGY-LOSS SPECTROSCOPY

被引:39
作者
LEAPMAN, RD [1 ]
ANDREWS, SB [1 ]
机构
[1] NINCDS, NEUROBIOL LAB, BETHESDA, MD 20892 USA
关键词
FIELD-EMISSION STEM; PARALLEL-DETECTION EELS; MASS MAPPING; ELEMENTAL MICROANALYSIS; DETECTION LIMITS; HEMOCYANIN; HEMOGLOBIN; TOBACCO MOSAIC VIRUS; RADIATION DAMAGE;
D O I
10.1111/j.1365-2818.1992.tb01482.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
The combination of scanning transmission electron microscopy (STEM) and parallel-detection energy-loss spectroscopy (EELS) was used to detect specific bound elements within macromolecules and macromolecular assemblies prepared by direct freezing. After cryotransferring and freeze-drying in situ, samples were re-cooled to liquid nitrogen temperature and low-dose (about 10(3) e/nm2) digital dark-field images were obtained with single-electron sensitivity using a beam energy of approximately 100 keV and a probe current of approximately 5 pA. These maps provided a means of characterizing the molecular weights of the structures at low dose. The probe current was subsequently increased to about 5 nA in order to perform elemental analysis. The 320 copper atoms in a keyhole limpet haemocyanin molecule (mol. wt = 8 MDa) were detected with a sensitivity of +/- 30 atoms in an acquisition time of 200 s. Phosphorus was detected in an approximately 10-nm length of single-stranded RNA contained in a tobacco mosaic virus particle (mol. wt = 130 kDa/nm) with a sensitivity of +/- 25 atoms. Near single-atom sensitivity was achieved for the detection of iron in one haemoglobin molecule (mol. wt = 65 kDa, containing four Fe atoms). Such detection limits are only feasible if special processing methods are employed, as is demonstrated by the use of the second-difference acquisition technique and multiple least-squares fitting of reference spectra. Moreover, an extremely high electron dose (about 10(10) e/nm2) is required resulting in mass loss that may be attributable to 'knock-on' radiation damage.
引用
收藏
页码:225 / 238
页数:14
相关论文
共 42 条
[1]  
ANDREWS SB, 1989, MICROBEAM ANAL, P85
[2]   UNCONVENTIONAL MODES FOR STEM IMAGING OF BIOLOGICAL STRUCTURES [J].
COLLIEX, C ;
JEANGUILLAUME, C ;
MORY, C .
JOURNAL OF ULTRASTRUCTURE RESEARCH, 1984, 88 (02) :177-206
[3]   RADIATION-DAMAGE DUE TO KNOCK-ON PROCESSES ON CARBON FOILS COOLED TO LIQUID-HELIUM TEMPERATURE [J].
DIETRICH, I ;
FOX, F ;
HEIDE, HG ;
KNAPEK, E ;
WEYL, R .
ULTRAMICROSCOPY, 1978, 3 (02) :185-189
[4]   ELECTRON ENERGY-LOSS SPECTROSCOPY AND CHEMICAL-CHANGE [J].
EGERTON, RF ;
CROZIER, PA ;
RICE, P .
ULTRAMICROSCOPY, 1987, 23 (3-4) :305-312
[5]   MASS MAPPING OF A PROTEIN COMPLEX WITH THE SCANNING-TRANSMISSION ELECTRON-MICROSCOPE [J].
ENGEL, A ;
BAUMEISTER, W ;
SAXTON, WO .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1982, 79 (13) :4050-4054
[6]   MOLECULAR-WEIGHT DETERMINATION BY SCANNING-TRANSMISSION ELECTRON-MICROSCOPY [J].
ENGEL, A .
ULTRAMICROSCOPY, 1978, 3 (03) :273-281
[7]  
HAIDER M, 1988, TUTORIALS INSTRUMENT, V1, P123
[8]  
HAMILTON MG, 1989, 47TH P ANN M EL MICR, P248
[9]  
HARRISON PM, 1980, IRON BIOCH MED, V2, P131
[10]   RECENT ASPECTS OF THE SUBUNIT ORGANIZATION AND DISSOCIATION OF HEMOCYANINS [J].
HERSKOVITS, TT .
COMPARATIVE BIOCHEMISTRY AND PHYSIOLOGY B-BIOCHEMISTRY & MOLECULAR BIOLOGY, 1988, 91 (04) :597-611