AN ULTRAHIGH-VACUUM SCANNING TUNNELING MICROSCOPE FOR USE AT VARIABLE-TEMPERATURE FROM 10-K TO 400-K

被引:38
作者
HORCH, S
ZEPPENFELD, P
DAVID, R
COMSA, G
机构
[1] Institut für Grenzflächenforschung und Vakuumphysik, KFA Forschungszentrum Jülich
关键词
D O I
10.1063/1.1145219
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report on the construction of an ultrahigh vacuum (UHV) scanning tunneling microscope (STM) specially designed for operation in the entire range of sample temperatures between 10 and 400 K. The sample is cooled by means of a liquid helium continuous-flow cryostat, while the supporting manipulator and the surrounding devices remain at room temperature. This allows rapid variation of the sample temperature. The standard instruments for surface preparation and analysis and the STM are contained in a single UHV chamber. By rotation of the manipulator the sample can be positioned in front of any of these instruments without changing the sample temperature. The performance of the microscope is demonstrated by two examples of images of xenon adsorbed on platinum(111) showing: (a) the evolution of the morphology of a submonolayer of xenon from adsorption at 17 K up to desorption at about 90 K and (b) atomically resolved images of the hexagonal incommensurate rotated phase for xenon at monolayer completion.
引用
收藏
页码:3204 / 3210
页数:7
相关论文
共 12 条
[1]   AN EASILY OPERABLE SCANNING TUNNELING MICROSCOPE [J].
BESOCKE, K .
SURFACE SCIENCE, 1987, 181 (1-2) :145-153
[2]  
COMSA G, 1992, SPRINGER SERIES SURF, V27, P243
[3]   HIGH-RESOLUTION HE-SCATTERING APPARATUS FOR GAS-SURFACE INTERACTION STUDIES [J].
DAVID, R ;
KERN, K ;
ZEPPENFELD, P ;
COMSA, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1986, 57 (11) :2771-2779
[4]   COARSE TIP DISTANCE ADJUSTMENT AND POSITIONER FOR A SCANNING TUNNELING MICROSCOPE [J].
FROHN, J ;
WOLF, JF ;
BESOCKE, K ;
TESKE, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (06) :1200-1201
[5]  
HORCH S, IN PRESS
[6]  
KERN K, 1988, J VAC SCI TECHNOL A, V3, P639
[7]  
MICHELY T, 1991, THESIS U BONN
[8]  
MICHELY T, 1989, J MICROSC-OXFORD, V152, pC10
[9]  
RODER H, 1991, THESIS U BONN
[10]   TECHNIQUES FOR ATTAINMENT, CONTROL, AND CALIBRATION OF CRYOGENIC TEMPERATURES AT SMALL SINGLE-CRYSTAL SAMPLES UNDER ULTRAHIGH-VACUUM [J].
SCHLICHTING, H ;
MENZEL, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07) :2013-2022