COMPARATIVE-ANALYSIS OF POWER-LINE INTERFERENCE BETWEEN 2-ELECTRODE OR 3-ELECTRODE BIOPOTENTIAL AMPLIFIERS

被引:19
作者
WOOD, DE
EWINS, DJ
BALACHANDRAN, W
机构
[1] UNIV SURREY,DEPT MECH ENGN,BIOMED ENGN GRP,GUILDFORD GU2 5XH,SURREY,ENGLAND
[2] UNIV SURREY,DEPT ELECTR & ELECT ENGN,GUILDFORD GU2 5XH,SURREY,ENGLAND
关键词
AC MAINS; BIOPOTENTIAL AMPLIFIERS; INTERFERENCE; REFERENCING; SPICE;
D O I
10.1007/BF02522948
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Differential input amplifiers are commonly used in the measurement of biopotential signals. This is because of their ability to reject common-mode signals, for example power-line interference. Rejection is further enhanced by referencing the subject to the instrumentation system. One such referencing technique uses a third subject electrode connected either to circuit common or to ground. However, these techniques are rarely explained or quantified in the literature. Four recording configurations are examined, isolated and non-isolated amplifiers with two and with three electrodes, and existing models are extended. From a brief assessment of power-line interference sources, it is concluded that only displacement currents induced into the leads and the body need be considered. The effect of these on the four configurations is analysed using SPICE. Results show that interference from displacement currents induced into the leads and into the body is dependent upon the recording configuration used and is significantly lower in three-electrode than in two-electrode amplifiers. Isolation also reduces levels of interference. It is concluded that the choice of referencing may affect interference levels on the biopotential signal significantly. Experimental and further analytical work is planned to confirm this.
引用
收藏
页码:63 / 68
页数:6
相关论文
共 15 条
[1]  
APPLE HP, 1976, POSTGRAD MED J, V52, P79
[2]   60-HZ INTERFERENCE IN ELECTROCARDIOGRAPHY [J].
HUHTA, JC ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (02) :91-101
[3]   MINIATURE SKIN-MOUNTED PREAMPLIFIER FOR MEASUREMENT OF SURFACE ELECTROMYOGRAPHIC POTENTIALS [J].
JOHNSON, SW ;
LYNN, PA ;
MILLER, JSG ;
REED, GAL .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1977, 15 (06) :710-711
[4]   AMPLIFICATION OF BIOSIGNALS BY BODY POTENTIAL DRIVING [J].
LEVKOV, CL ;
DASKALOV, IK .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1982, 20 (02) :248-250
[5]   MAINS INTERFERENCE IN ECG RECORDING [J].
MARSHALL, I ;
NEILSON, JM .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1984, 8 (04) :177-180
[8]   SKIN IMPEDANCE FROM 1 HZ TO 1 MHZ [J].
ROSELL, J ;
COLOMINAS, J ;
RIU, P ;
PALLASARENY, R ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (08) :649-651
[9]  
SVETZ P, 1975, ELECTRON DES, V16, P68
[10]   GROUND-FREE ECG RECORDING WITH 2 ELECTRODES [J].
THAKOR, NV ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (12) :699-704