FURTHER-STUDIES ON THE EFFECTS OF MAGNETIC-RESONANCE-IMAGING FIELDS ON MIDDLE-EAR IMPLANTS

被引:40
作者
APPLEBAUM, EL
VALVASSORI, GE
机构
[1] BERWYN MAGNET IMAGING CTR,BERWYN,IL
[2] UNIV ILLINOIS,COLL MED,DEPT RADIOL,CHICAGO,IL 60612
关键词
magnetic resonance imaging; middle ear implants; stapedectomy prostheses;
D O I
10.1177/000348949009901009
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
We investigated the effects of magnetic resonance imaging (MRI) fields on 21 stapedectomy prostheses and other middle ear implants and two different receiver-stimulator modules from 22-channel cochlear implants. None of the middle ear implants was displaced by the magnetic field, except for one platinum–stainless steel stapedectomy piston. Magnetism was not induced in any of the middle ear implants subjected to prolonged exposure in the MRI scanner. We conclude that MRI could pose a hazard to patients who have had stapedectomy using certain platinum–stainless steel piston prostheses and to patients with cochlear implants. Magnetic resonance imaging should pose no hazard to patients who have had the other middle ear implants reported on in this and our previous investigation. © 1990, SAGE Publications. All rights reserved.
引用
收藏
页码:801 / 804
页数:4
相关论文
共 12 条
  • [1] APPLEBAUM EL, 1985, ARCH OTOLARYNGOL, V111, P820
  • [2] BOESIGER RB, 1988, MAGNET RESON MED, V7, P255
  • [3] OPERATIONAL SAFETY ISSUES IN MRI
    GANGAROSA, RE
    MINNIS, JE
    NOBBE, J
    PRASCHAN, D
    GENBERG, RW
    [J]. MAGNETIC RESONANCE IMAGING, 1987, 5 (04) : 287 - 292
  • [4] HEPFNER ST, 1985, NEW ENGL J MED, V313, P387
  • [5] MIDDLE-EAR PROSTHESIS - SIGNIFICANCE IN MAGNETIC-RESONANCE-IMAGING
    LEON, JA
    GABRIELE, OF
    [J]. MAGNETIC RESONANCE IMAGING, 1987, 5 (05) : 405 - 406
  • [6] LOAKMAN RW, 1985, RADIOLOGY, V157, P711
  • [7] THE EFFECT OF MAGNETIC-RESONANCE IMAGING ON METAL SPINE IMPLANTS
    LYONS, CJ
    BETZ, RR
    MESGARZADEH, M
    REVESZ, G
    BONAKDARPOUR, A
    CLANCY, M
    [J]. SPINE, 1989, 14 (07) : 670 - 672
  • [8] THE EFFECT OF NUCLEAR-MAGNETIC-RESONANCE IMAGING ON METALLIC MIDDLE-EAR PROSTHESES
    MATTUCCI, KF
    SETZEN, M
    HYMAN, R
    CHATURVEDI, G
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1986, 94 (04) : 441 - 443
  • [9] MR IMAGING OF METALLIC IMPLANTS AND MATERIALS - A COMPILATION OF THE LITERATURE
    SHELLOCK, FG
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (04) : 811 - 814
  • [10] SHELLOCK FG, 1989, JAMA-J AM MED ASSOC, V261, P3412