HYBRID COMPUTERIZED FLUOROSCOPY TECHNIQUE FOR NON-INVASIVE CARDIOVASCULAR IMAGING

被引:44
作者
ERGUN, DL
MISTRETTA, CA
KRUGER, RA
RIEDERER, SJ
SHAW, CG
CARBONE, DP
机构
关键词
D O I
10.1148/132.3.739
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The excellent linearity of digital image storage and retrieval permits hybrid analog-digital subtraction to extend the spatial resolution of two previously developed algorithms which employed entirely digital apparatus. A low resolution, time-integrated preinjection digital mask image is reconverted to analog form and subtracted from live analog video images of iodine administered by peripheral intravenous injection to produce a high resolution display of the cardiovascular system with contrast ten times greater than conventional fluoroscopy. Preliminary studies in dogs are compared with images obtained with the digital subtraction algorithms.