GAS-STREAM HEATER TO CONTROL LOCAL SKIN TEMPERATURE UNDER EVAPORATIVE SWEAT CAPSULES

被引:3
作者
KRANING, KK [1 ]
SCHIBLI, EG [1 ]
MOMII, ST [1 ]
MERRICK, EA [1 ]
机构
[1] UNIV WASHINGTON,DEPT ELECT ENGN,SEATTLE,WA 98195
关键词
D O I
10.1152/jappl.1978.44.6.981
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A compact 4-W gas-stream heater for use with ventilated sweating capsules is described. The heating element is constructed from two modified 75-Ω power resistors connected in parallel and contained in the capsule gas inlet tube. Heater power is controlled with integrated circuits which compare the output of a thermocouple, measuring either capsule skin or air temperature, with a predetermined target temperature. The unit responds fairly slowly to sudden large changes in target temperature, but once a steady state is achieved, maintains target temperature levels despite changes in evaporative heat loss. The device is well suited for studies in which local skin temperature or capsule air temperature must be elevated and clamped while the effects of other local stimuli or central thermoregulatory activity are studied, and in studies where the skin temperature effect on transepidermal water loss must be controlled.
引用
收藏
页码:981 / 985
页数:5
相关论文
共 7 条
[1]   MEASUREMENT OF EVAPORATIVE WATER LOSS BY A THERMAL CONDUCTIVITY CELL [J].
ADAMS, T ;
FUNKHOUSER, GE ;
KENDALL, WW .
JOURNAL OF APPLIED PHYSIOLOGY, 1963, 18 (06) :1291-&
[2]   USE OF DEW-POINT DETECTION FOR QUANTITATIVE MEASUREMENT OF SWEATING RATE [J].
BRENGELMANN, GL ;
MCKEAG, M ;
ROWELL, LB .
JOURNAL OF APPLIED PHYSIOLOGY, 1975, 39 (03) :498-500
[3]   CONTINUOUS RECORDING OF SWEATING RATE BY RESISTANCE HYGROMETRY [J].
BULLARD, RW .
JOURNAL OF APPLIED PHYSIOLOGY, 1962, 17 (04) :735-737
[4]  
FOSTER KG, 1967, J PHYSIOL-LONDON, V191, pP1
[5]  
GASSELT HRM, 1963, DERMATOLOGIA, V127, P225
[6]   MECHANISM OF ENHANCEMENT OF ECCRINE SWEATING BY LOCALIZED HEATING [J].
MACINTYRE, BA ;
BULLARD, RW ;
BANERJEE, M ;
ELIZONDO, R .
JOURNAL OF APPLIED PHYSIOLOGY, 1968, 25 (03) :255-+